ARTICLE IN PRESS
Biomaterials ] (]]]]) ]]]–]]]
www.elsevier.com/locate/biomaterials
Review
Medical applications of poly(styrene-block-isobutylene-block-styrene)
(‘‘SIBS’’)
Leonard Pinchuka,Ã, Gregory J. Wilsonb, James J. Barryc, Richard T. Schoephoersterd,
Jean-Marie Parele,f, Joseph P. Kennedyg
a
Innovia LLC and InnFocus LLC, 12415 SW 136 Ave Unit 3, Miami, FL 33186, USA
b
The Hospital for Sick Children, Toronto, Ontario, Canada
c
Boston Scientific Corporation, Natick, MA, USA
d
Florida International University, Miami, FL , USA
e
Ophthalmic Biophysics Center, Bascom Palmer Eye Institute, University of Miami Miller School of Medicine, Miami, FL, USA
f
University of Liege, CHU Sart-Tilman, Liege, Belgium
g
The University of Akron, Akron, OH, USA
Received 13 July 2007; accepted 30 September 2007
Abstract
Poly(Styrene-block-IsoButylene-block-Styrene) (‘‘SIBS’’) is a biostable thermoplastic elastomer with physical properties that overlap
silicone rubber and polyurethane. Initial data collected with SIBS stent-grafts and coatings on metallic stents demonstrate
hemocompatibility, biocompatibility and long-term stability in contact with metal. SIBS has been used successfully as the carrier for
a drug-eluting coronary stent; specifically Boston Scientific’s TAXUSs stent, and its uses are being investigated for ophthalmic implants
to treat glaucoma, synthetic heart valves to possibly replace tissue valves and other applications. At present, researchers developing
medical devices utilizing SIBS have found the following: (1) SIBS does not substantially activate platelets in the vascular system; (2)
polymorphonuclear leukocytes in large numbers are not commonly observed around SIBS implants in the vascular system or in
subcutaneous implants or in the eye; (3) myofibroblasts, scarring and encapsulation are not clinically significant with SIBS implanted in
the eye; (4) embrittlement has not been observed in any implant location; (5) calcification within the polymer has not been observed; and
(6) degradation has not been observed in any living system to date. Some deficiencies of SIBS that need to be addressed include creep
deformation in certain load-bearing applications and certain sterilization requirements. The reason for the excellent biocompatibility of
SIBS may be due to the inertness of SIBS and lack of cleavable moieties that could be chemotactic towards phagocytes.
r 2007 Elsevier Ltd. All rights reserved.
Keywords: Biostable; Poly(styrene-b-isobutylene-b-styrene); SIBS; Glaucoma; Polyurethane
Contents
1.
2.
3.
4.
5.
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
The degradation of polyurethanes and events leading to SIBS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
The SIBS hypothesis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
The synthesis and properties of SIBS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
SIBS in medicine . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5.1. Stent-grafts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5.2. SIBS as the drug carrier in the TAXUSs drug-eluting coronary stent . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
ÃCorresponding author. Tel.: +1 305 378 2651x284; fax: +1 305 378 2652.
E-mail address: len@innovia-llc.com (L. Pinchuk).
0142-9612/$ - see front matter r 2007 Elsevier Ltd. All rights reserved.
doi:10.1016/j.biomaterials.2007.09.041
Please cite this article as: Pinchuk L, et al. Medical applications of poly(styrene-block-isobutylene-block-styrene).... Biomaterials (2007), doi:10.1016/
j.biomaterials.2007.09.041
2
2
3
5
6
6
7
ARTICLE IN PRESS
L. Pinchuk et al. / Biomaterials ] (]]]]) ]]]–]]]
2
6.
5.3. SIBS in ophthalmology: the MIDI-Tube glaucoma shunt . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5.4. SIBS for a synthetic trileaflet aortic valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Conclusions and future work . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Acknowledgments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
References . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1. Introduction
Poly(Styrene-block-IsoButylene-block-Styrene) (‘‘SIBS’’)
was relatively unknown in medicine prior to the introduction of Boston Scientific Corporation’s (BSC) (Natick,
MA) Drug Eluting TAXUSs Coronary Stent in 2002. This
medical device has significantly reduced the incidence of
coronary bypass procedures and associated morbidities.
SIBS is a thermoplastic elastomer whose physical properties overlap both the silicone rubbers and polyurethanes.
SIBS is oxidatively, hydrolytically and enzymatically stable
over its lifespan in the body and is, therefore, biostable
with a relatively low foreign body reaction.
The development of SIBS for medical devices evolved
from deficiencies encountered with the long-term in vivo
performance of polyurethanes; specifically, their degradation and concomitant inflammatory and fibrotic reactions.
These deficiencies limited the use of polyurethanes for longterm load-bearing implant applications and for applications in contact with metal, as metal ions contribute to
polyurethane degradation by oxidative pathways. It is
important to understand these nuances to appreciate the
significance of the polymer chemistry of SIBS; a brief
review of the use of polyurethanes in the body will
therefore also be presented.
The inertness of SIBS has enabled novel medical devices,
such as the aforementioned TAXUSs Stent, a minimally
invasive glaucoma drainage tube implant for treating
glaucoma (Miami-Innfocus Drainage Implant—‘‘MIDITube’’) and a synthetic trileaflet aortic valve to possibly
replace tissue and mechanical valves. Several other medical
devices made from SIBS are in early stages of development,
e.g., spinal implants, and may be commercialized over the
next few years.
2. The degradation of polyurethanes and events leading to
SIBS
In the early 1980s, Medtronic Corporation (Minneapolis, MN) introduced the polyether urethane-insulated
pacemaker lead. The polyurethane was manufactured by
Dow Chemical and sold as Pellethanes 2363 80A, a
thermoplastic aromatic polyether urethane. By the mid1980s, reports describing surface pitting and cracking of
the polyether urethane lead insulator began appearing in
the literature. In the early 1990s, the degradation of these
polyurethanes became widespread and represented the
limiting factor in the development of novel medical devices
using microfilamentous scaffolds requiring load-bearing
.9
11
11
11
12
properties, such as compliant vascular grafts (see Fig. 1A).
A review of the state-of-the-art of polyurethanes in the
mid-1990s was presented by Pinchuk [1]. Contrary to what
had been reported in the literature [2], these microfilamentous polyurethanes did not need to be stressed to exhibit
biodegradation. Simple relaxed subcutaneous implants of
these microporous high surface area materials in animals
showed significant biodegradation in less than 4 weeks.
This implant model was used to investigate a variety of
polyurethanes with soft segments including polydimethylsiloxane (PDMS) diols and fluorinated PDMS diols. In
addition, surface treatment of conventional polyurethanes,
such as with plasma polymerized polytetrafluoroethylene
(PTFE) and surface-grafted PDMS, were also investigated
[3,4]. It was found that surface treatment or surface
grafting, did not protect the polyurethane from biodegradation as, in the case of PTFE coatings, they cracked due
to compliance mismatch thereby exposing the underlying
polyurethane to the surrounding bodily fluids. Although
degradation was delayed for months by the use of surfacegrafted PDMS, this coating was sufficiently oxygen
permeable to enable oxidation of the underlying polyurethane. A variety of polyurethanes with various soft
segments containing PDMS and surface modifying agents
have appeared in the literature over the last decade;
however, these materials exhibited similar biodegradation
[5–11].
The degradation mechanism of polyether urethanes was
elucidated by Anderson’s group [12,13] at Case Western
Reserve University (Cleveland, Ohio). They found that the
carbon alpha to the ether of the polyether soft segment was
oxidized to ester either by superoxide produced by
polymorphonuclear leucocytes (PMNs) and the like, or
by metal ion contact of the polyurethane, as occurs on the
inside of pacemaker lead insulators. Subsequent hydrolysis
of the ester cleaves the macromolecule, and in the presence
of flexion, cracks develop. Realizing that the ether groups
were vulnerable, Pinchuk introduced more biostable
polycarbonate urethanes for implant applications which
were initially commercialized under the trade name:
CorethaneTM (Corvita Corp., Miami, FL (acquired by
Boston Scientific Corporation, Natick, MA)) [14,15]. The
polycarbonate urethane patents were transferred to the
Polymer Technology Group (Berkeley, CA) in 1996 and
various formulations of these polyurethanes are currently
being marketed under the trade name ‘‘Bionate’’.
The motivation to develop biostable polyurethanes
stemmed from the need for elastomeric materials to
provide compliant vascular grafts as well as deformable
Please cite this article as: Pinchuk L, et al. Medical applications of poly(styrene-block-isobutylene-block-styrene).... Biomaterials (2007), doi:10.1016/
j.biomaterials.2007.09.041
ARTICLE IN PRESS
L. Pinchuk et al. / Biomaterials ] (]]]]) ]]]–]]]
3
Fig. 1. (A) Non-implanted spun tubular meshwork vascular graft. (B) SEM of the graft showing 10 mm diameter polycarbonate urethane fibers as
compared to a human hair (shown on the diagonal). (C) Meshwork at 2 years explant showing a virtually intact fibrous structure. (D) Numerous fibers of
the same graft demonstrating surface cracking.
stent-graft liners [16–18]. The improved biostability of
polycarbonate urethanes was confirmed by Stoke’s group
at Medtronic using the ‘‘Stokes Test’’, in which a tube of
the material is stretched over a dumbbell-shaped mandril
and exposed to oxidizing and hydrolyzing chemicals, or is
implanted in the body for a predetermined time [19].
Materials that are readily susceptible to oxidation and
hydrolysis crack in this model; significantly, the polycarbonate urethanes did not crack.
Although polycarbonate urethanes demonstrated superior biostability relative to the polyether and polyester
urethanes, they too eventually exhibited biodegradation as
manifested by surface cracking. Fig. 1 shows a scanning
electron micrograph (SEM) of a spun-polycarbonate
urethane aorto-iliac vascular graft explanted from a dog
at 2 years. Fig. 1A shows the non-implanted spun tubular
meshwork; Fig. 1B is an SEM of the graft showing the
polycarbonate urethane microfibers (diameter approximately 10 mm) in comparison to a human hair (Fig. 1B)
which is 10 times larger in diameter. Fig. 1C is an explant
of the polycarbonate urethane meshwork at 2 years
showing the fibrous structure virtually intact, whereas
Fig. 1D shows numerous fibers of the same graft
demonstrating surface cracking. The fractures were most
noticeable in areas with large numbers of macrophages on
histology (not shown).
Importantly, Wilson’s group (The Hospital for Sick
Children, Toronto, Ontario) also observed that these
degrading implants attracted a plethora of PMNs, especially during the early weeks of implantation (Fig. 2A).
Further, the cleaner the polycarbonate urethane (less
extractables, washed surfaces), the more intense the
inflammation. Further observations were the attraction of
macrophages, foreign body giant cells and the phagocytosis
of small ‘‘chunks’’ of polyurethane (Figs. 2B and C).
Lastly, it was also observed upon careful examination that
crack formation in the microfilamentous grafts occurred in
areas of the graft as early as 1 month after implantation.
In summary, polyurethanes exhibit degradation with
time with signs of the problem occurring within weeks of
implantation. Degradation is due to oxidation, most likely
by superoxide produced by phagocytes (‘‘scavenger cells’’);
the more degradation, the greater number of scavenger
cells that migrate to the site to engulf and therefore
sequester the eluting material. Because phagocytes release
substances which stimulate fibrosis, the eluting material
becomes encapsulated; that is, walled off. By the mid1990s, it became apparent that new polymers were needed
for medical devices that required long-term implantation
for load-bearing applications. This was particularly essential in microporous embodiments as well as in products
utilizing metals, such as stents, stent-grafts, drug-eluting
stents and pacer leads.
3. The SIBS hypothesis
An analysis of the degradation mechanism of polyurethanes combined with an understanding of organic
chemistry principles led to the hypothesis that the longterm stability of a polymeric material in living tissue can be
achieved when both the polymeric backbone and pendant
groups are devoid of unprotected ester, amide, ether,
Please cite this article as: Pinchuk L, et al. Medical applications of poly(styrene-block-isobutylene-block-styrene).... Biomaterials (2007), doi:10.1016/
j.biomaterials.2007.09.041
ARTICLE IN PRESS
4
L. Pinchuk et al. / Biomaterials ] (]]]]) ]]]–]]]
Fig. 2. The arrows point to: (A) polymorphonuclear leukocytes around an 8-week subcutaneous implant of a polycarbonate urethane in a dog
(unpublished work courtesy of G. Wilson, The Hospital for Sick Children, Toronoto Ontario Canada); (B) foreign body giant cells in a 12-week
polyurethane implant intrastromal in the cornea of a rabbit eye; (C) phagocytosis of a polyurethane by foreign body giant cells in a 21-week intrastromal
implantation in the cornea of a rabbit eye (B and C are provided courtesy of M. Fukuda et al.; Kinki University School of Medicine, Osaka, Japan).
carbamate, urea, or any other groups that are prone to
oxidation, hydrolysis, or enzymatic cleavage. Further, the
degradation of polyethylene acetabular joint liners that
produce in-chain unsaturation and crosslinking [20–23],
that dominated the literature for the last two decades,
suggest that secondary carbon-containing polymers such as
polyethylene, and secondary-and-tertiary carbon-containing polymers such as polypropylene [24–26], are also to be
avoided as double bond formation leads to embrittlement
and stress cracking. It must also be kept in mind that,
depending on the polymerization processes used, polyethylene always contains a number of methyl (or higher
alkyl) branches, and therefore contains a corresponding
number of oxidizable tertiary hydrogens ($CH(CH3)
CH2$). These irregularities are not protected within the
polyethylene crystallites, and, particularly when exposed
on surfaces, are accessible to oxidative attack by the body’s
defense systems.
Consequently, it was hypothesized that an ideal polymer
for implant application should contain only oxidatively,
hydrolytically and enzymatically stable alternating secondary-and-quaternary carbons in the backbone, and equally
stable primary carbons as pendant groups. The basic
structures of this nature are those comprised of polyisobutylene (PIB) shown on the left of Fig. 3. The absence of
cleavable side groups in PIB, in contrast to potentially
hydrolyzable ester groups in methacrylate polymers, e.g., the
methoxy group in poly(methylmethacrylate), shown on the
right of Fig. 3, which contains a similar alternating
secondary–quaternary carbon backbone, should provide a
polymer with less biodegradation. Further, if this hypothesis
is correct, the less biodegradation, the less inflammation.
Secondary
Carbon
Quaternary
Carbon
CH3
CH2
CH3
CH2
C
C
CH3
Polyisobutylene
C
O = OCH3
Poly(methylmethacrylate)
Fig. 3. Schematic of polyisobutylene with its alternating secondary and
quaternary carbons and lack of labile pendent groups as compared to
poly(methylmethacrylate) with its ester side group.
PIB, an inert non-vulcanizable rubber used in many
industrial applications (i.e., tackifiers, adhesives, sealants,
thickening agents, viscosity enhancers, various additives,
chewing gum, etc.) can be obtained easily and inexpensively by the cationic polymerization of isobutylene.
However, PIB cannot be used in applications where shape
retention is essential because it is not crosslinked. A very
close relative to PIB is butyl rubber, a commercially
available copolymer of $98% isobutylene and $2%
isoprene, in which the few but critically important isoprene
units –CH2–C(CH3) ¼ CH–CH2– provide vulcanizability,
that is shape retention. However, butyl rubber is also
unsuitable for implantation in living tissue as: (1) it
contains oxidatively vulnerable unsaturations, and (2) it
can be converted into a shape-retaining rubber only by
vulcanization under harsh, biologically unacceptable,
conditions with crosslinkers and additives that are generally not tolerated in the body.
Please cite this article as: Pinchuk L, et al. Medical applications of poly(styrene-block-isobutylene-block-styrene).... Biomaterials (2007), doi:10.1016/
j.biomaterials.2007.09.041
ARTICLE IN PRESS
L. Pinchuk et al. / Biomaterials ] (]]]]) ]]]–]]]
The search for PIB-based thermoplastic elastomers, i.e.,
for elastomers that contains PIB rubbery segments
covalently linked to readily thermally- and/or solutionprocessible glassy segments, led the lead author to
Kennedy’s laboratory at The University of Akron where
such polymers were already synthesized [27]. Kennedy’s
patents protecting the triblock copolymer SIBS were
licensed by Corvita Corporation (Miami, FL, Acquired
by BSC, Natick, MA) and strengthened by additional
patents covering applications in the medical implant arena
[28–30].
The triblock SIBS, was used for the initial studies of
biocompatibility and biostability. Fig. 4 shows the
molecular structure of SIBS in which soft PIB rubbery
chains are held together by hard glassy polystyrene
domains. Fig. 5 shows a cartoon of the synthesis and
architecture of this thermoplastic elastomer. Dangling
chains are absent and all the PIB segments contribute
to the load-bearing capacity of the network. SIBS is
a self-assembled physically crosslinked PIB, and thus
thermo- and solution-formable. Furthermore, because it
is soluble in various non-polar solvents it can be spraycoated or solvent cast to deliver soft strong coherent
films.
CH3
H
CH3
C CH2
C CH2
N
CH3
H
CH2 C
CH3
M
H 2C C
N
M
Fig. 4. Schematic of poly(styrene-block-isobutylene-block-styrene)
(‘‘SIBS’’), where N/(M+N) for biomedical application is generally
0.05–0.50. X is the residue of the hindered dicumyl ether initiator
(see Fig. 6).
I
Ionization
G G- +
Initiation
+G
iBu
I
G+
PIB
St
pSt
PIB
Glassy Rubbery
PIB
Propagation
PIB
+G-
Triblock Formation
pSt
Rubbery Glassy
Phase separated
thermoplastic
elastomer
Fig. 5. Cartoon of the synthesis stages of SIBS, showing ionization,
initiation, propagation and block formation.
5
CH3
CH3O
C
CH3
CH3
OCH3
C
CH 3
Fig. 6. 5-Tert-butyl-1,3-bis(1-methoxy-l-methylethyl)-benzene, also called
5-tert-butyl-1,3-dicumyl ether (MW 278), AKA hindered dicumyl ether
(HDCE).
4. The synthesis and properties of SIBS
SIBS is synthesized by the living cationic polymerization
technique developed by Kennedy’s team at The University
of Akron. Living cationic polymerization, a seminal
discovery in polymer science, lead not only to SIBS but
also to many novel compositions useful for a variety of
industrial and medical applications [27].
The synthesis of SIBS begins with a bifunctional
initiator, which becomes part of the polymer. The preferred
initiator is 5-tert-butyl-1,3-bis(1-methoxy-l-methylethyl)benzene (for brevity’s sake ‘‘hindered dicumyl ether’’,
abbreviated HDCE). HDCE is not commercially available
and is custom synthesized by Innovia LLC (Miami, FL)
according to methods developed by Kennedy et al. [31,32].
In brief, SIBS is prepared in two steps in one pot: First
isobutylene is polymerized by a HDCE/TiCl4 initiating
system in a methyl chloride/hexanes solvent system in the
presence of a proton trap under a blanket of dry nitrogen
at À80 1C. When the central PIB block reaches the desired
molecular weight, styrene is added and the polymerization
is continued until the outer polystyrene blocks also reach a
predetermined length. The process is terminated by the
addition of methanol. Fig. 7 outlines the synthesis of SIBS.
Table 1 presents typical properties of SIBS. The
molecular weight of the triblock is controlled by reaction
conditions, mainly by the ratio of monomers/initiator. The
hardness of SIBS can be varied by the amount of styrene
employed. Fig. 8 shows a plot of Shore hardness in both
the A scale and D scale as a function of mole percent
polystyrene in SIBS.
The excellent oxidative stability of SIBS can be demonstrated by submerging a swatch of SIBS in boiling
concentrated (65%) nitric acid for 30 min. Whereas other
elastomers used for implant applications, such as silicone
rubber and polycarbonate urethane, severely embrittle or
are completely destroyed within a few minutes, SIBS
remains relatively unscathed and stable under these harsh
conditions [28]. Silicone rubber (PDMS) is well-known to
degrade by strong acids and strong bases [33].
SIBS can be injection and compression molded, as well
as extruded and solvent cast from non-polar solvents such
as methylcyclohexane, cyclopentane, toluene and tetrahydrofuran. Furthermore, components made from SIBS
can be solvent-bonded with these non-polar solvents.
Please cite this article as: Pinchuk L, et al. Medical applications of poly(styrene-block-isobutylene-block-styrene).... Biomaterials (2007), doi:10.1016/
j.biomaterials.2007.09.041
ARTICLE IN PRESS
L. Pinchuk et al. / Biomaterials ] (]]]]) ]]]–]]]
6
CH2O
OCH3
CH3
mH2C
CH3
IB
HDCE
Cl
m-1
m-1
TiCl4
-80°C
C
_
_
m-1
m-1
PlB
Styrene
1)nH2C
CH
2) MeOH
PlB
Cl
Cl
n
m
m
Cl
n
Poly(styrene-block-isobutylene-block-styrene)
Fig. 7. Reaction scheme for the synthesis of SIBS using living cationic polymerization.
Table 1
Typical physical properties of SIBS triblock polymers
Shore hardness
Mole percent styrene
Ultimate tensile strength (psi)
Ultimate tensile strength (MPa)
Ultimate elongation (%)
Index of refraction
Water absorption (g/m2 at 24 h)
Weight average molecular weight
Polydispersity (Mw/Mn)
30A–60D
5–50
2000–5000
10–30
300–1100
1.525–1.535
0.2–0.3
60,000–150,000
1.2–2.1
not gamma-ray sterilizable; (6) the cost of synthesis and
purification of SIBS is, at the present production rate,
relatively high, and; (7) due to intellectual property
constraints, there is no source for implant-grade material,
other than small amounts, that may or may not be
available from Innovia LLC.
Shore Hardness
120
100
Shore A
80
60
Shore D
40
5. SIBS in medicine
20
5.1. Stent-grafts
0
0
10
20
30
40
50
60
Mole Percent Styrene
Fig. 8. Plot of mole percent styrene in SIBS versus shore hardness
(courtesy of Yonghua Zhou and John B. Martin (Innovia LLC, Miami,
FL)).
The downside of the thermoplastic nature of SIBS is that
(1) it is susceptible to stress cracking in the presence of
organic solvents; (2) it has poor creep properties and
therefore requires fiber reinforcement for certain loadbearing applications (see heart valves below); (3) the nonpolar nature of SIBS does not allow sites for hydrogen
bonding, as in polyurethanes, which limits its ultimate
tensile strength to less than that of a dry polyurethane;
(4) it has poor gas permeability which renders it more
cumbersome to sterilize with ethylene oxide as all surfaces
must be exposed to the gas (this is in contrast to silicone
rubber which is permeable to ethylene oxide); (5) SIBS is
The first two embodiments of SIBS to be tested for longterm implant were: (1) a microporous liner on a stent;
i.e., as a stent-graft, and (2) approximately 1 mm diameter
monofilaments. The stent-grafts were constructed by
rotating a braided wire mesh stent on a mandril while
spraying it with SIBS dissolved in tetrahydrofuran. The
SIBS filaments were formed in mid-air as the tetrahydrofuran flashed off thereby depositing a porous tubular mat.
Animal studies were initiated with a protocol that required
explants of the monofilaments at 1 and 3 months, and
explants of the porous stent-graft at 6, 12 and 24 months.
The monofilament data demonstrated absence of biodegradation by scanning electron microscopy examination
and by tensile strength measurement [34,35]. The 6, 12 and
24-month explants of the stent-graft liners showed intact
structures with no fiber cracking or biodegradation by
scanning electron microscopy (tensile strength data were
not acquired because the liners were attached to stent
Please cite this article as: Pinchuk L, et al. Medical applications of poly(styrene-block-isobutylene-block-styrene).... Biomaterials (2007), doi:10.1016/
j.biomaterials.2007.09.041
ARTICLE IN PRESS
L. Pinchuk et al. / Biomaterials ] (]]]]) ]]]–]]]
A
B
7
stent wires
SIBS
liner
SIBS liner
Vessel
Lumen
Explanted Consortemer CEG
2 years follow-up, Study C-33 Pilot
T.S. #2516 Lt. lliac, HPSx20
C
Explanted Consortemer CEG
2 years follow-up, Study C-33 Pilot
T.S. #2516 Lt. lliac, HPSx100
D
E
10µm 400x
Fig. 9. (A) Cross-section of a 6 mm diameter stent-graft, made with sprayed SIBS fibers explanted after 24 months in the iliac artery of a dog (HPS stain
(20 Â )). (B) The same graft as (A) at 100 Â magnification showing excellent tissue ingrowth. (C) A pre-implant SEM of the sprayed SIBS microfibers.
(D) and (E) are explants at 6 and 24 months, respectively, showing no fiber cracking (C–E are from Ref. [35]).
struts). Figs. 9A and B show the histology of the explanted
stent-graft at 2 years demonstrating excellent tissue
integration and wide patency. Figs. 9C–E show approximately 5 mm diameter fibers, before implant and at both 6
months and at 2 years explant indicating no cracking.
Molecular weight constancy of these polymers was corroborated using gel permeation chromatography (GPC) of
1-year explants from coronary stents by Steckel et al. [36]
from Barry’s group at Boston Scientific. It is noteworthy
that the SIBS coating was in intimate contact with the
stainless steel stent and remained unaffected by metal ion
oxidation.
It was also observed throughout the implant study by
Wilson’s group, as well as by Boston Scientific researchers,
and in many other SIBS explants in both animals and
humans, that there were usually very few, if any, PMNs
around SIBS, which is in contrast to the significant
accumulation of PMNs observed around implanted polyurethanes. The smaller recruitment of PMNs suggests that
the biostable nature of SIBS may play a significant role in
controlling the foreign body reaction.
5.2. SIBS as the drug carrier in the TAXUSs drug-eluting
coronary stent
Boston Scientific Corporation (BSC—Natick, MA)
scientists began evaluating SIBS as a drug carrier in their
drug-eluting coronary stent program. BSC was actively
researching the use of paclitaxel, delivered in a controlled
manner from a stent to minimize the process of in stent
restenosis (re-occlusion of the coronary artery after stent
implantation) which occurred in approximately 35% of all
patients receiving a coronary stent. Paclitaxel, at appropriate concentrations, functions by promoting the formation of stable microtubules in cells and reduces
proliferation and migration of smooth muscle cells in the
arterial wall. It is through these proliferative processes that
the lumen of the artery can narrow. Two comprehensive
reviews of paclitaxel and its mechanism of interaction with
smooth muscle cells on a stent, as well as the preclinical
and clinical trials of the TAXUSs drug-eluting stent, are
provided by Kamath et al. [37,38] from Barry’s group. The
present publication briefly reviews the importance of SIBS
in this drug-eluting application. Boston Scientific scientists
were attempting to release paclitaxel from a polycarbonate
urethane coated onto their stent. Several in vivo studies
showed a consistent inflammatory response characterized
by PMN infiltration around polyurethane-coated stent
struts. Fig. 10A shows a cross-section of a 2-month explant
from a porcine coronary artery implanted with a noncoated-bare-metal coronary stent (BMS) showing the
expected widely patent lumen and paucity of inflammation.
However, when the stent struts were coated with polycarbonate urethane, it elicited a very strong inflammatory
Please cite this article as: Pinchuk L, et al. Medical applications of poly(styrene-block-isobutylene-block-styrene).... Biomaterials (2007), doi:10.1016/
j.biomaterials.2007.09.041
ARTICLE IN PRESS
8
L. Pinchuk et al. / Biomaterials ] (]]]]) ]]]–]]]
Fig. 10. Cross-section of porcine coronary arteries with stent explants at various time points: (A) BMS (non-coated stent) at 2 months; (B) a
polycarbonate urethane-coated stent with significant inflammation and hyperproliferation (2 months); (C) a high magnification of B showing the presence
of PMNs (2 months). (D) BMS (3 months); (E) and (F) are stents coated with SIBS (90 and 180 days, respectively), showing wide patency. (BSC in
collaboration with Elazer Edelman and Campbell Rogers, MIT [polyurethane studies, Rob Schwartz, Mayo Clinic] and Greg Wilson, Hospital for Sick
Children—Toronto, Ontario.)
reaction and ultimately lumen loss (Fig. 10B). Magnification of the strut area (Fig. 10C) showed significant
inflammation with infiltration by PMNs and macrophages.
The hyperplastic response secondary to inflammation
resulted in luminal narrowing and occlusion of the artery.
The aggressive cellular reaction around the polyurethanecoated coronary stent was essentially identical to that
observed around the degrading polyurethane vascular
grafts shown in Fig. 1. This was a surprising observation
for the BSC team given the widely reported successful use
of polyurethanes in numerous medical applications. For a
drug-eluting stent utilizing a polymer carrier to be
successful, a highly desirable characteristic is that the
polymer should not elicit a biological response worse than
that of the BMS. BSC researchers subsequently began
evaluating SIBS as a means of minimizing the inflammatory reaction.
SIBS was tested on a coronary stent in the porcine
coronary model and it was confirmed that SIBS demonstrated a biological response similar to the BMS.
Figs. 10D–F show a BMS stent following 90 days
implantation as well as a SIBS-coated stent at 90 and 180
days post-operatively, respectively. These experiments
clearly showed that patency in these 3 mm diameter
coronary arteries could be maintained with SIBS-coated
stents. In addition Kamath reported that SIBS-coated
stents showed platelet adhesion and activation at a level
comparable to that of the BMS in the absence and presence
of paclitaxel, and this observation held true for periods
greater than 1 year in their challenging porcine coronary
model. Further testing by Barry’s group with stents coated
with SIBS/paclitaxel combinations revealed, after a few
dosing iterations, that paclitaxel released in a controlled
manner from SIBS could mitigate the restenosis of
coronary arteries in animals [37–42]. Extensive clinical
trials have confirmed these same findings in humans [43].
Fig. 11 shows an atomic force micrograph (AFM) as well
as a transmission electron micrograph (TEM) of SIBS/
paclitaxel coated on a coronary stent. Although paclitaxel
is lipophilic, it is insoluble in the hydrophobic SIBS and
tends to disperse into small islands within the polymer
matrix. Accordingly, after an initial burst from the surface
following implantation, most of the paclitaxel remains in
the polymer. Work on blends of SIBS with other polymers,
such as poly(styrene-r-maleic anhydride) [44] and poly(alkyl methacrylate-b-isobutylene-b-alkyl methacrylate) [45]
are ongoing to improve control of the release kinetics of
paclitaxel as well as other drugs from the bulk of the drug
carrier. Initial clinical trials of the TAXUSs stent were
conducted by Grube and Bullesfeld [46] (Heart-Center
Siegburg, Germany) who reported on BSC’s TAXUS-I
safety trial (paclitaxel/SIBS-coated NIR stent (Tel Aviv,
Israel)). At 6 months follow-up, there were 0% restenosis
in the TAXUS arm versus 10% in the BMS control group.
Please cite this article as: Pinchuk L, et al. Medical applications of poly(styrene-block-isobutylene-block-styrene).... Biomaterials (2007), doi:10.1016/
j.biomaterials.2007.09.041
ARTICLE IN PRESS
L. Pinchuk et al. / Biomaterials ] (]]]]) ]]]–]]]
9
Fig. 11. AFM and TEM of SIBS showing islands of 8.8% w/w paclitaxel (see arrows) that are insoluble in the SIBS matrix (white rectangles) (Courtesy of
Ranade et al., from BSC).
In addition, there were no incidences of thrombotic events.
These studies combined with the TAXUS II study [47]
helped support filings with the United States Food and
Drug Administration resulting in the ultimate release of the
TAXUSs Stent to the US market.
Several other studies of the TAXUSs Stent were
published following its initial launch and are summarized
in Kamath’s review articles [37,38]. The TAXUSs Stent
continues to demonstrate the beneficial long-term durability and efficacy of a drug-eluting stent in preventing
restenosis as well as its safety out to 4 years follow-up from
the time of this writing [43]. It is important to recognize
that other groups have attempted to place paclitaxel on a
stent with limited success in reducing restenosis [48,49].
This exemplifies the importance of the polymer carrier, in
this case SIBS, as a critical component for the success of
drug-eluting stents for treating coronary artery disease.
More recently, it has been reported that all drug-eluting
stents available in the US are indicating a slight increase in
late stent thrombosis that appears to be primarily
attributed to patients prematurely discontinuing their
systemic anti-coagulation therapy [50]. However, a recent
FDA sponsored expert panel on this topic overwhelmingly
supported the benefits of drug-eluting stents over the risk
of late stent thrombosis in patients who are indicated for a
drug-eluting stent [51].
The drug-eluting stent is the most recent treatment of
coronary artery disease over the last 30 years. In a field that
evolved from coronary bypass surgery, to balloon angioplasty with up to 60% restenosis rates, to bare-metal stents
with up to 40% restenosis, to the drug-eluting stent—this
technology offers a significant improvement of quality of
life for hundreds of thousands of patients.
5.3. SIBS in ophthalmology: the MIDI-Tube glaucoma
shunt
Work performed by Parel and his colleagues (The University of Miami’s, Miller School of Medicine, Bascom
Palmer Eye Institute, Miami FL) in conjunction with
scientists at InnFocus LLC (an Innovia spin-off, Miami,
FL), showed that SIBS disks implanted in the stroma of
the cornea (Fig. 12), and in the sub-Tenon space in the eye,
are significantly less irritating and inflammatory than
similarly shaped PDMS controls [52,53]. It was discovered during this first set of implants in rabbits that
SIBS barely encapsulates in the eye as compared to PDMS,
which demonstrates significant encapsulation and concomitant angiogenesis. The first application leveraging
this finding was a glaucoma drainage tube, called the
‘‘MIDI-Tube’’, first described by Acosta et al. [54] from
Parel’s group. The MIDI-Tube, illustrated in Fig. 12,
is an 11 mm long, 250 mm diameter SIBS microtube
with a small tab (fin) located half-way along the length of
the tube to prevent migration of the tube into the anterior
chamber.
The lumen of the tube, determined by the HagenPoiseuille equation [55], is between 60 and 100 mm in
diameter to prevent hypotony (deflation of the eye with
subsequent collapse of the anterior chamber) without the
need for an intrinsic pressure relief valve. The tube is
introduced into the anterior chamber from a needle tract
located 2 mm posterior to the limbus that is tunneled under
the limbus and exits in the anterior chamber. As illustrated
in Fig. 12, the proximal end of the MIDI bisects the angle
between the iris and the cornea with the tip of the tube
located 2 mm into the anterior chamber. The distal end of
the tube rests in a subconjunctival/Tenon flap created
posterior to the limbus, extending to the equator of the eye.
A shallow bleb (a blister-like structure) forms in the
conjunctival flap as aqueous humor drains from the
anterior chamber. Fluid from the bleb is absorbed into
the venous system of the eye or through the conjunctiva
into the tear ducts, depending upon the health of the tissue
surrounding the device. This shunting of fluid from the
high-pressure interior of the eye to the near atmospheric
pressure of the bleb, effectively lowers the pressure in
the eye.
Please cite this article as: Pinchuk L, et al. Medical applications of poly(styrene-block-isobutylene-block-styrene).... Biomaterials (2007), doi:10.1016/
j.biomaterials.2007.09.041
ARTICLE IN PRESS
L. Pinchuk et al. / Biomaterials ] (]]]]) ]]]–]]]
10
Canal of
Schlemm
Conjunctiva
MIDI-Tube
Fin
Intra-Stromal Disk Implant
Corneal
Stroma
Limbus
a
ne
r
Co
Trabecular
Network
Tenons
Sub-Tenon
SIBS Disk
Implant
IRIS
eb
Bl
ra
le
Sc
Lens
Ciliary
Muscles
Retina
Ciliary Body
Ligaments of
Zonules
Fig. 12. Schematic of the eye showing the sites for the intrastromal SIBS disk implant and the sub-Tenon SIBS disk implant. Also shown is the MIDITube, which conducts aqueous humor from the anterior chamber to a bleb formed under the conjunctiva/Tenons as a means of reducing pressure in the
eye to treat glaucoma.
The MIDI-Tube was tested in 41 New Zealand White
Rabbits; 22 with a 70 mm, 6 with a 100 mm and 4 with a
150 mm lumen diameter tube and 9 with PDMS tube
controls (OD, 600 mm; ID, 300 mm; length, 19 mm). Followup at various intervals up to 14 months showed no
inflammation, migration or extrusion of the MIDI-Tube.
Flow patency in the SIBS group was confirmed in all cases
by fluorescein (0.01% in saline) injection into the anterior
chamber with subsequent drainage through the tube into
the bleb. There were no incidences of hypotony or flat
chambers at post-operative day 1. Slit-lamp and optical
coherence tomography (OCT) exams showed clearance of
the proximal entrance and no clogging along the tube, a
normal anterior chamber and a quiet ocular surface. Light
microscopy and immunostaining against collagen IV,
macrophages and a-smooth muscle actin demonstrated a
distinct absence of myofibroblasts and inflammatory cells
with minimal scarring in the MIDI group. In contrast, the
silicone rubber tubes demonstrated patency in only 2/9
devices at 6 months and immunostaining showed myofibroblasts and significant capsular formation. An independent 6-month biocompatibility study of the MIDI-Tube
(without PDMS controls), performed at NAMSA (North
American Science Associates, Inc., Toledo, OH) in 10
rabbits using good laboratory practices (GLPs), yielded
similar results with SIBS tubes, showing only one migration of a device due to an iatrogenically damaged tab
[52,56–58].
It is interesting to speculate why silicone rubber leads to
significant encapsulation in both the initial disk implants
and in the glaucoma tube studies within the subconjunctiva
of eye, whereas SIBS does not, keeping in mind that tissue
interfaces methyl groups in both polymers. PDMS, which
is synthesized by ring opening polymerization of octamethylcyclotetrasiloxane, may elute unreacted cyclics into
adjacent tissue and may attract PMNs. In addition, silicone
rubber is unstable at both high and low pH [33], and acids
produced by PMNs, macrophages and the like may cleave
moieties which may be chemotactic towards additional
phagocytes. In contrast, SIBS is stable at extreme pH
ranges.
SIBS has also been used successfully as a scleral buckle
in rabbits for the possible reattachment of detached retinas
[59]. The advantage of SIBS for this application is that
SIBS does not significantly encapsulate and is therefore
thinner with less chance of restricting eye motion.
Furthermore, it is removable as compared to silicone
rubber, which tends to form large capsules and remain in
the patient for life.
SIBS is also a candidate for use as an intraocular lens
(IOL) or anterior chamber lens as it has a high refractive
index of 1.525–1.535 at all hardnesses. In addition, it does
not have a yield point at durometers less than Shore 50A
and can be folded or rolled without scratching or surface
cracking and therefore can be introduced into the lens
capsule through a cannula less than 2 mm in diameter.
Further, due to the high flex fatigue life of SIBS (see heart
valve section), it may be of use as an accommodating IOL
to treat presbyopia. Accommodating IOLs may enable
post-cataract patients to see both far and near without
wearing spectacles. Lenses, both with and without haptics,
have been cryogenically machined from SIBS and there
is no reason why such lenses could not be injection
or compression molded. Pilot implantations of two
IOLs in two rabbits at the Bascom Palmer Eye Institute’s
Ophthalmic Biophysics Center, by Viviana Fernandez,
showed excellent biocompatibility. What remains to
be seen is whether the creep behavior of SIBS will force
it to take an undesirable shape in the lens sac. Lastly,
it has been hypothesized that the lack of cleavable
side groups from SIBS may provide a lens with less
chronic corneal endothelial damage as compared to
Please cite this article as: Pinchuk L, et al. Medical applications of poly(styrene-block-isobutylene-block-styrene).... Biomaterials (2007), doi:10.1016/
j.biomaterials.2007.09.041
ARTICLE IN PRESS
L. Pinchuk et al. / Biomaterials ] (]]]]) ]]]–]]]
Fig. 13. The all-synthetic flow-through heart leaflet valve for replacement
of the aortic valve. The leaflets are comprised of a composite of PET and
SIBS. The valve stent is of a harder durometer SIBS (courtesy of Siobhain
Lowe, Florida International University).
conventional lenses that show continual loss of endothelial
cells with time [60]. This hypothesis has yet to be
substantiated.
5.4. SIBS for a synthetic trileaflet aortic valve
Schoephoerster’s group at Florida International University (Miami, FL) demonstrated that SIBS has a high flex
fatigue life and when reinforced with PET fibers or fabric,
exhibits minimal creep deformation. Unreinforced SIBS,
like most other thermoplastics, will creep. Creep resistance,
combined with excellent hemocompatibility and biostability, suggests that SIBS may be suitable for use in a throughflow heart leaflet valve [61–66]. A composite heart valve
leaflet has been designed (Fig. 13) with low modulus SIBS
(7–9 mol% styrene), reinforced with a knitted polyester
fabric, and attached to a SIBS frame of significantly higher
modulus (35–40 mol% styrene). Animal implants and longterm in vitro flex fatigue studies are ongoing (fatigue lives
are currently in excess of 400 million cycles, which
approximates 10 years of operation in a human). It is
anticipated that the inertness of SIBS will prevent
degradation and intrinsic calcification which has been the
bane of other polymers, especially polyurethane, which
have been tested extensively for this application.
6. Conclusions and future work
SIBS is a biostable thermoplastic elastomer with properties ideal for certain medical implant applications. Initial
data collected with SIBS liners on stent-grafts and coatings
on stents demonstrate hemocompatibility, biocompatibility
and long-term stability in contact with metal. SIBS has
been used successfully in drug-eluting coronary stents and
11
its uses are being investigated for ophthalmic implants,
synthetic heart valves and other applications. At the
present, researchers developing medical devices utilizing
SIBS have found the following: (1) SIBS does not
substantially activate platelets in the vascular system;
(2) PMNs in large numbers are not commonly observed
around SIBS implants in the vascular system or in
subcutaneous implants or in the eye; (3) myofibroblasts,
scarring and encapsulation are not clinically significant
with SIBS implanted in the eye; (4) embrittlement has not
been observed in any implant location; (5) calcification
within the polymer has not been observed; (6) degradation
has not been observed in any living system to date;
(7) creep deformation needs to be addressed in certain
load-bearing applications; and (8), care must be taken
when sterilizing SIBS with ethylene oxide to ensure all
surfaces are well-exposed to the gas. The reason for the
excellent biocompatibility of SIBS may be due to the
inertness of SIBS and lack of cleavable moieties that could
be chemotactic towards inflammatory cells. This inertness
distinguishes SIBS from other implantable polymers such
as the acrylates, methacrylates, polyesters, polyethers,
polyamides and polyurethanes that contain ester, ether,
amide or carbamate groups, which may slowly hydrolyze,
oxidize or cleave side groups which may be chemotactic
towards phagocytes. An abundance of these scavenger cells
can lead to thick capsules and/or other undesirable
sequelae.
PIB-based polymers are continuously evolving. Kennedy’s and Puskas’s groups at The University of Akron;
Faust’s laboratory at the University of Massachusetts
(Lowell, MA); Innovia and its Affiliates and BSC are
working on advanced chemistries that include injectable
polymers that polymerize in the body, as well as
hydrophilic and enhanced hydrophobic designs for controlled drug delivery. In addition, tougher, and more
lubricious materials are being developed for orthopedic
and spinal applications to replace the failing polyethylenes
[23] and polyurethanes [1] which are prone to oxidation by
body fluids. It is anticipated that PIB-based polymers will
play a major role in many new medical devices in the
future.
Acknowledgments
The six co-authors of this review are the team leaders
from multiple academic institutions and commercial
entities who have been responsible for bringing SIBS
research and medical product development to its present
state. The six teams collectively comprise many engineers,
chemists, materials scientists, physicians and academic
trainees to whom we are grateful. In regard specifically to
manuscript preparation, the authors wish to thank Marcia
Orozco for her help in organizing this manuscript,
Yonghua Zhou and John Martin for providing the data
on percent styrene versus hardness of SIBS and Kalpana
Kalmath for help in TAXUSs research.
Please cite this article as: Pinchuk L, et al. Medical applications of poly(styrene-block-isobutylene-block-styrene).... Biomaterials (2007), doi:10.1016/
j.biomaterials.2007.09.041
ARTICLE IN PRESS
12
L. Pinchuk et al. / Biomaterials ] (]]]]) ]]]–]]]
References
[1] Pinchuk L. A review of the biostability and carcinogenicity of
polyurethanes in medicine and the new generation of biostable
polyurethanes. J Biomater Sci Polym Ed 1994;6:225–67.
[2] Zhao Q, McNally AK, Rubin KR, et al. Human plasma a2macroglobulin promotes in vitro oxidative stress cracking of
pellethane 2363-80A: in vivo and in vitro correlations. J Biomed
Mater Res 1993;27:379–89.
[3] Pinchuk L, Crack prevention of implanted prostheses; grafting
a siloxane elastomer onto biocompatible surface. US patent no
4851009, 1989.
[4] Pinchuk L, Esquivel MC, Martin Jr. JB. Polyurethane/silicone
composites for long-term implant in the human body. In: Proceedings
of the society of plastic engineerings ANTEC, Montreal, May 1991.
p. 1802–4.
[5] Adhikari R, Gunatillake PA, Maijs GF, Martin DJ, Poole-Warren
LA, Schindhelm K. Synthesis and biostability of low modulus
siloxane-based polyurethanes. Transactions of the society for
biomaterials: sixth world biomaterials congress, Minneapolis, part
II, 2000, 812pp.
[6] Ward RS, Tian Y, Chen Q, Nguyen C, Serpilio J. High strength,
optically clear, silicone-urethane thermoplastics for biomedical use:
bulk properties. Transactions of the society for biomaterials: sixth
world biomaterials congress, Minneapolis, part I, 2000, 431pp.
[7] Christenson EM, Anderson JM, et al. the effect of silicone on the
biostability of poly(ether urethanes). Transactions of the society for
biomaterials: 28th annual meeting, Tampa, Florida, April 24 to 27,
2002.
[8] Ebert M, Ward R, Anderson J, Tain Y, McVenes R, Stokes K.
Chronic Environmental Stress Cracking Resistance of Polyurethanes
with Polyether and Polysiloxane Soft Segments. Society of Biomaterials: 29th Annual Meeting Transactions, 2003. p. 535.
[9] Ebert M, Ward R, Anderson J, Tain Y, McVenes R, Stokes K.
Screening of polyurethanes with polyether and polysiloxane soft
segments for resistance to environmental stress cracking. society of
biomaterials: 29th annual meeting transactions, 2003, 542pp.
[10] Ebert M, Ward R, Anderson J, Tain Y, McVenes R, Stokes K.
Chronic. Environmental stress cracking resistance of polyetherurethanes with fluorocarbon surface modifying endgroups. society of
biomaterials: 29th annual meeting transactions, 2003, 538pp.
[11] Ebert M, Ward R, Anderson J, Tain Y, McVenes R, Stokes K.
Chronic environmental stress cracking resistance of polyetherurethanes with polyethyleneoxide surface modifying endgroups. society
of biomaterials: 29th annual meeting transactions, 2003, 536pp.
[12] Wu Y, Sellitti C, Anderson JM, Hiltner A, Lodoen GA, Payet CR.
An FTIR-ATR investigation of in vivo poly(ether urethane)
degradation. J Appl Polym Sci 1992;46:201.
[13] Anderson JM, Hiltner A, Zhao QA, Wu Y, Renier M, Schubert M,
et al. Cell/polymer interactions in the biodegradation of polyurethanes. In: Vert M, Feijen J, Albertson A, Scott G, Chiellini E,
editors. Biodegradable polymers and plastics. Cambridge, England:
Royal Society of Chemistry; 1992. p. 122–36.
[14] Pinchuk, L. Crack-resistant polycarbonate urethane polymer prostheses. US patent no 5133742, 1992.
[15] Pinchuk L, Kato YP, Eckstein ML, Wilson GJ, MacGregor DC.
Polycarbonate urethanes as elastomeric materials for long-term
implant applications. Transactions of the 19th annual meeting of
the society for biomaterials. Birmingham, AL, 1993, 22pp.
[16] MacGregor DC, Wilson GJ, Klement P, Weber BA, Binnington AG,
Pinchuk L. A new compliant composite elastomer/dacron small
caliber vascular prosthesis. Transactions of the 17th annual meeting
of the society for biomaterials, vol. 14. Scottsdale, AZ, 1991, 145pp.
[17] Wilson GJ, MacGregor DC, Klement P, Dereume JP, Weber BA,
Binnington AG, et al. The composite corethane/dacron vascular
prosthesis: canine in vivo evaluation of 4 mm diameter grafts with 1
year follow-up. Trans Am Soc Artif Organs 1991;37:475–6.
[18] Wilson GJ, Klement P, Kato YP, Omar S, Wilson P, Bridgeman J, et
al. Healing of a self-expanding bifurcated vascular graft for
abdominal aortic aneurysm repair in dogs. Transactions of the 44th
annual conference asaio/cvst, april 1998.
[19] Zhou Q, Cass-Bejar I, Urbanski P, Stokes K. Glass wool-H2O2/
CoCl2 test system for in vitro evaluation of biodegradable stress
cracking in polyurethane elastomers. J Biomed Mater Res 1995;
29:467–75.
[20] Jasty M, Goetz DD, Bragdon CR, Lee KR, Hanson AE, Elder JR, et
al. Wear of polyethylene acetabular components in total hip
arthroplasty. An analysis of one hundred and twenty-eight components retrieved at autopsy or revision operations. J Bone Jt Surg
1997;79:349–58.
[21] Kurtz SM. The UHMWPE handbook: ultra-high molecular weight
polyethylene in total joint replacement. New York: Academic Press;
2004.
[22] Wright TM, Goodman SB. Implant wear. Rosemont, IL: American
Academy of Orthopedic Surgeons; 2001.
[23] Kurtz SM, Hozack W J, Purtill JJ, Marcolongo M, Kraay M J,
Goldberg VM, et al. Otto Aufranc award paper: significance of in
vivo degradation for polyethylene in total hip arthroplasty. Clin
Orthop Relat Res 2006;453:47–57.
[24] Deacon J, Apple DJ. Further studies of IOL materials. In:
Proceedings of the 11th annual meeting of the society for
biomaterials. San Diego, CA: University of Alabama, Birmingham;
1985. 205pp.
[25] Goldberg EP, Yalon M, Stacholy J, Kurobe N, Longo WE, Osborn
DC, et al. Biocompatibility of intraocular lens polymers. In:
Proceedings of the 11th annual meeting of the society for
biomaterials. San Diego, CA: University of Alabama, Birmingham;
1985. 208pp.
[26] Apple DJ, Mamalis N, Brady SE, Loftfield K, Kavka-Van Norman
D, Olson R J. Biocompatibility of implant materials: a review and
scanning electron microscopy study. J Am Intraocul Implant Soc
1984;10(1):53–66.
[27] Kennedy JP. From thermoplastic elastomers to designed biomaterials. J Polym Sci [A1] 2005;43:2863–951.
[28] Pinchuk L. Biostable elastomeric polymers having quaternary
carbons. US patent no 5741331, 1998.
[29] Pinchuk L. Method of implanting biostable elastomeric polymers
having quaternary carbons. US patent no 6102939, 2000.
[30] Pinchuk L. Method of device having biostable elastomeric polymers
having quaternary carbons. US patent no 6197240, 2001.
[31] Wang B, Mishra MK, Kennedy JP. Living carbocationic polymerization XII. Telechelic polyisobutylenes by a sterically hindered
bifunctional initiator. Polym Bull 1987;17:205–11.
[32] Mishra MK, Kennedy JP. Living carbocationic polymerization VIII.
Telechelic polyisobutylenes by the MeO(CH2)2C-p-C5H4-C(CH3)2
Ome/BCl3 initiating system. Polym Bull 1987;17:7–13.
[33] Liebhafsky HA. Silicones under the monogram: a story of industrial
research. New York: Wiley; 1978 [p. 281–282].
[34] Pinchuk L, Khan IJ, Martin JB, Bridgeman J, Wilson GJ, Glass J, et
al. A new family of thermoplastic elastomers for ultra-long term
implant based upon a backbone of alternating quaternary and
secondary carbons. Transactions of the 44th annual conference
ASAIO/CVST, 1998.
[35] Pinchuk L, Khan IJ, Martin JB, Wilson GJ. Polyisobutylene-based
thermoplastic elastomers for ultra long term implant applications.
Transactions of the society of biomaterials, sixth world biomaterials
congress transactions, Hawaii, 2000, 1452pp.
[36] Steckel M, Richard RE, Straumann P, Barry JJ. In vivo biostability
of SIBS controlled release polymer in the taxustm drug-eluting
coronary stents. Transactions of biomaterials in regenerative
medicine: the advent of combination products, October 16–18,
Philadelphia, PA, 2004, 42pp.
[37] Kamath KR, Barry JJ, Miller KM. The TAXUSTM drug eluting
stent: a new paradigm in controlled drug delivery. Adv Drug Deliv
Rev 2006;58:412–36.
Please cite this article as: Pinchuk L, et al. Medical applications of poly(styrene-block-isobutylene-block-styrene).... Biomaterials (2007), doi:10.1016/
j.biomaterials.2007.09.041
ARTICLE IN PRESS
L. Pinchuk et al. / Biomaterials ] (]]]]) ]]]–]]]
[38] Kamath KR, Barry JJ. The utilization of antiproliferative and
antimigratory compounds for the prevention of restenosis. In:
Kipshidze N, Fareed J, Moses J, Serruys P, editors. Textbook of
Interventional cardiovascular pharmacology. Informa Healthcare,
London, Boca Raton, FL: Taylor & Francis; 2007. p. 299–314.
[39] Pinchuk L, Nott S, Schwarz M, Kamath K. Drug delivery
compositions and medical devices containing block copolymer. US
patent no 6545097, 2003.
[40] Richard RE, Barry JJ, Kamath K, Miller K, Renade SV, Schwarz M.
Polymer technology for the controlled release of paclitaxel from a
vascular compatible matrix—the TAXUSTM express paclitaxeleluting coronary stent. Transactions of biomaterials in regenerative
medicine: the advent of combination products, October 16–18,
Philadelphia, PA, 2004. p. 9–10.
[41] Ranade SV, Miller KM, Richard RE, Chan AK, Allen MJ, Helmus
MN. Physical characterization of controlled release of Paclitaxel from
the TAXUSTM Express2TM drug-eluting stent. J Biomed Mater Res
A 2004;71:625–34.
[42] Ranade SV, Richard RE, Helmus MN. Styrenic block copolymers for
biomaterial and drug delivery applications. Acta Biomater 2005;
1(1):137–44.
[43] Stone GW, Moses JW, Ellis SG, Schofer J, Dawkins KD, Morice
MC, et al. Safety and efficacy of sirolimus- and paclitaxel-eluting
coronary stents. N Engl J Med 2007;356:998–1008.
[44] Schwartz MC, Ranade SV, Richard RE. Controlled delivery of
paclitaxel from stent coatings containing styrene maleic anhydride
polymer blends. In: 31st Annual meeting of the controlled release
society, 12–16 June, Honolulu, Hawaii, USA, 2004.
[45] Cho JC, Cheng G, Feng D, Faust R, Richard R, Schwarz M, et al.
Synthesis and characterization, propertues, and drug release of
poly(alkyl methacrylate-b-isobutylene-b-alkyl methacrylate). Biomacromolecules 2006;7(11):2997–3007.
[46] Grube E, Bullesfeld L. Initial experience with paclitaxel-coated stents.
J Interv Cardiol 2002;15:471–5.
[47] Gregg W, Stone GW. TAXUS clinical trials: TCT recap. Cath Lab
Digest 2004;12(12):30–2.
[48] Grube E, Lansky A, Hauptmann KE, Di Mario C, Di Sciascio G,
Colombo A, et al. High-dose 7-hexanoltaxol-eluting stent with
polymer sleeves for coronary revascularization: one year results from
the SCORE randomized trial. J Am Coll Cardiol 2004;44(7):1368–72.
[49] Silber S. Paclitaxel-eltuting stents: are they all equal? An analysis of
randomized controlled trials in de novo lesions of 3319 patients. J
Interv Cardiol 2003;16(6):485–90.
[50] Dave RM, Sharma SK. The state of coronary drug-eluting stent use.
current issues in DES care. Card Interventions Today 2007;1(1):
33–43.
[51] Christy P, editor-in-chief. The state of coronary drug-eluting stent
use: FDA panel addresses late stent thrombosis in DES, Card
Interventions Today, 2007;1(1):44–52.
[52] Acosta AC, Fernandez V, Lamar P, Orozco M, Dubovy S, Stoiber J,
Pinchuk L, et al. Ocular biocompatibility of quatromer (polystyrenepolystyrene triblock polymers) for glaucoma applications. Invest
Ophthalmol Vis Sci 2004;45: E-abstract 2929.
13
[53] Parel J-M, Stoiber J, Fernandez V, Acosta AC, Lamar PD, Dubovy
S, et al. Ocular and orbital biocompatibility of a novel implant
material (SIBS): a preliminary study, in preparation.
[54] Acosta AC, Espana EM, Yamamoto H, Davis S, Pinchuk L,
Weber BA, et al. A newly designed glaucoma drainage implant
made of poly(styrene-b-isobutylene-b-styrene) biocompatibility
and function in normal rabbit eyes. Arch Ophthalmol 2006;124:
1742–9.
[55] Pinchuk L, Fantes F, Zhou Y, Martin JB, Parel J-M. The
BPEI-innfocus glaucoma drainage implant—the MIDI: concept,
physics and pressure control. Invest Ophthalmol Vis Sci 2006;47:
E-Abstract 36.
[56] Acosta AC, Espana EM, Orozco M, Yamamoto H, Pinchuk L,
Dubovy S, et al. A novel glaucoma implant: a comparative SIBS vs
PDMS study in normal rabbits. Invest Ophthalmol Vis Sci 2006;47:
E-Abstract 3545.
[57] Fantes F, Acosta AC, Carraway J, Siddiq F, Pinchuk L, Weber BA,
et al. An independent GLP evaluation of a new glaucoma drain, the
MIDI. Invest Ophthalmol Vis Sci 2006;47: E-Abstract 3547.
[58] Orozco M, Acosta AC, Espana EM, Davis S, Weber BA, Pinchuk L,
et al. New styrene-b-isobutylene-b-styrene (SIBS) glaucoma drainage
implant. Ophthalmic technologies XVI. Proceedings of the SPIE, vol.
6138, 2006, p. 177–84.
[59] Aly M, Yamamoto H, Espana EM, Orozco MA, Aguilar M, Arrieta
E, et al. A novel material for scleral buckle procedure: an
experimental comparative pilot study, in preparation.
[60] Perez-Santoja JJ, Alio JL, Jimenez-Alfaro I, Zato MA. Surgical
correction of severe myopia with an angle-supported phakic
intraocular lens. J Cataract Refract Surg 2000;26(9):1288–302.
[61] Schoephoerster RT, Gallocher S, Pinchuk L, Kasjanov VA. A novel
trileaflet synthetic heart valve. Adv Bioeng: ASME BED 2001;51:
225–6.
[62] Yin W, Gallocher S, Pinchuk L, Schoephoerster R, Jesty J, Bluestein
D. Comparative measurements of flow induced platelet activation in
a trileaflet polymeric valve and a St. Jude medical bileaflet mechanical
heart valve. Transactions of the Society For Biomaterials, Philadelphia, October, 2004.
[63] Gallocher SL, Aguirre A, Kasyanov V, Pinchuk L, Schoephoerster
RT. A novel polymer with potential use in a trileaflet heart valve. J
Biomed Mater Res B Appl Biomater 2006;79(2):325–34.
[64] Yin W, Gallocher S, Pinchuk L, Schoephoerster R, Jesty J, Bluestein
D. Flow induced platelet activation in a st. jude mechanical heart
valve, a trileaflet polymeric heart valve and a st jude tissue valve. Artif
Organs 2000;29(10):826–31.
[65] Wang Q, Gallocher SL, Mathison M., Kato Y, Pinchuk L,
Schoephoerster RT. In vivo pre-clinical evaluation of a novel
polymeric aortic valve. In: Proceedings of the 2006 BMES annual
fall meeting, October 11–14, Chicago, IL, 2006.
[66] Gallocher S, Wang Q, Mathison M, Kato Y, Pinchuk L,
Schoephoerster RT. Feasibility analysis of the safety and durability
of a polymer trileaflet heart valve. In: Proceedings of the 2006
summer Bioengineering conference, June 21–25, Amelia Island
plantation, Amelia Island, FL, 2006.
Please cite this article as: Pinchuk L, et al. Medical applications of poly(styrene-block-isobutylene-block-styrene).... Biomaterials (2007), doi:10.1016/
j.biomaterials.2007.09.041