ADHESIVE SEALANT
BIOMATERIALS
Technical Bulletin
Tensile Strength and Burst Pressure
Comparison of TissuePatchTM3 with other
commonly used Surgical Sealants
Lesley Shea, Analytical Scientist
Tissuemed Ltd.
______________________
_______________
Summary
The aim of this study was to compare the
physical properties (tensile strength and burst
pressure) of TissuePatch3 with commercially
available
comparator
products,
namely
TisseelTM(Baxter), TachoSil®(Nycomed) and
PleuraSealTM(Covidien).
The tensile strength of TissuePatch3 was
statistically greater than that of TachoSil (ttest, p=0.01), under both wet and dry
conditions. The tensile strength of PleuraSeal
and Tisseel could not be assessed, due to
problems during sample preparation.
The mean burst pressure of Tisseel was
significantly lower than that of TissuePatch3 (ttest, p=0.01). There was no significant
difference between the mean burst pressures
of the other products, probably due to the high
standard deviations observed.
Background
The aim of this study was to compare the
physical properties of TissuePatch3 with those
of the following products: Tisseel, TachoSil,
PleuraSeal, these being commonly used
products and representative of the categories
of product used as surgical sealants.
Methods
Tensile Strength Testing
Tensile strength testing of TissuePatch3 and
TachoSil (wet and dry) was conducted
according to Tissuemed Standard Operating
Procedure
5.12
Tensile
Testing
of
TissuePatch3.
Preparation of PleuraSeal and Tisseel samples
involved applying each sealant to a mould and
allowing the test strips to set.
Burst Pressure Testing
Burst Pressure testing was conducted in
accordance with ASTM F 2392-04 Standard
Test Method for Burst Pressure Strength of
Surgical Sealants.
Collagen films (4 mm diameter) were washed
in distilled water, and then soaked in fresh
distilled water for five minutes before being
stored in saline solution until use (for a
maximum of one hour).
A collagen film was removed from saline and
patted dry with blue roll. A 3 mm diameter hole
was punched in the centre of the collagen film.
A 15 mm diameter disc of TissuePatch3 or
TachoSil was applied over the 3 mm diameter
hole within the collagen film, with pressure
being applied for 30 seconds.
For applying Tisseel and PleuraSeal, an
approximately 1 mm thick mould was placed
over the collagen film, so that the 15 mm hole
of the mould was centrally placed over the 3
mm hole within the collagen film. The sealants
were prepared according to the manufacturer’s
instructions, and applied so that the hole of the
mould was filled. Tisseel and PleuraSeal were
left to set for 30 seconds before the mould was
carefully removed.
The collagen film was placed on the base of
the burst pressure test rig. The burst pressure
test rig top was put into position before being
secured with the o-rings and wing nuts. The
pressure gauge was zeroed and the peristaltic
pump was switched on at a flow rate of 2
ml/minute. The burst pressure of each sample
was recorded.
Results and Discussion
Tensile Strength Testing
When PleuraSeal was applied to the mould, it
flowed underneath the top layer. After five
minutes, the top part of the mould was
removed. The PleuraSeal had not fully set and
was observed to be ‘jelly-like’. The PleuraSeal
was left to fully set in the bottom of the mould
but could not be cut into strips. The sealant
tore, forming pieces that were not suitable for
testing.
Problems were also experienced during the
application of PleuraSeal to the mould, with the
application mechanism becoming blocked on
several occasions.
As a result the tensile strength testing of
PleuraSeal (and Tisseel) according to the test
protocol was abandoned.
The tensile strength of TachoSil was
statistically lower (under both dry and wet
conditions), when compared with the tensile
strength of TissuePatch3 (t-test, p=0.01).
Manufactured by: Tissuemed Ltd.•5 Killingbeck Drive•Leeds•LS14 6UK•United Kingdom
www.tissuemed.com•Tel +44(0)1132000500•Fax +44(0)1132407343•©Tissuemed 2009
TB17/01
Table 1 – Tensile Strength of TachoSil and TissuePatch3
TENSILE STRENGTH (kPa)
TachoSil
TissuePatch3 (TP3-015-08)
Dry
Wet
Dry
Wet
62.51
16.53
24197.42
1294.75
(± 18.66)
(± 5.90)
(± 7814.30)
(± 171.24)
Mean
Table 2 – Burst Pressure of PleuraSeal, TachoSil, Tisseel and TissuePatch3
Test #
Burst Pressure (cm H20)
PleuraSeal
TachoSil
Tisseel
TissuePatch3
Mean
78.0 ± 53.8
55.4 ± 59.5
8.1 ± 6.2
140.6 ± 58.2
Note: Due to the fast-setting nature of PleuraSeal only a low number of samples could be tested.
Burst Pressure Testing
Table 2 shows the comparative performance of
the products tested.
The mean burst pressure of Tisseel was
statistically lower than that of TissuePatch3 (ttest, p=0.01). There was no significant
difference between the mean burst pressures
of any of the other sealants, which can
probably be attributed to the relatively high
standard deviations observed and low sample
sizes of PleuraSeal and TachoSil tests.
Mean Burst Pressure (cm H 2 0)
250
200
150
100
50
0
PleuraSeal
T achoSil
T isSeel
T issuePatch3
-50
Sealant
Figure 1 Mean Burst Pressure of PleuraSeal, TachoSil, Tisseel and TissuePatch3
Conclusions
This in vitro assessment of the clinically
relevant
physical
characteristics
of
TissuePatch3 suggests that it provides acute
performance significantly better in some
parameters and equivalent in others when
compared
with
commercially
available
comparators.
Tensile Strength
The tensile strength of TissuePatch3 was
statistically greater (t-test, p=0.01) than the
tensile strength of TachoSil, when tested under
both dry and wet conditions.
Tensile testing of the liquid sealants according
to the established model rendered no
meaningful results.
Consequently new
methods for assessing tensile strength of these
polymerising liquids may need to be
established, although the fundamental lack of
strength in these polymer matrices may
ultimately mean such tests yield no useful data.
Burst Pressure
The mean burst pressure of Tisseel was
statistically lower (t-test, p=0.01) than that of
TissuePatch3 (TP3-015-08). There were no
statistical differences in the mean burst
pressures of the other surgical sealants,
probably due to the high standard deviations
observed.
References
Data on file at Tissuemed
ASTM method F 2392-04 Standard Test Method
for Burst Strength of Surgical Sealants.
Note:
Tisseel is a trademark of Baxter International
Inc PleuraSeal is a trademarks of Covidien AG
or its affiliate
TachoSil is a registered trademark of Nycomed
International Management GmbH
Tissuemed and TissuePatch are trademarks of
Tissuemed Ltd.
Manufactured by: Tissuemed Ltd.•5 Killingbeck Drive•Leeds•LS14 6UK•United Kingdom
www.tissuemed.com•Tel +44(0)1132000500•Fax +44(0)1132407343•©Tissuemed 2009