Biomaterials 21 (2000) 2335}2346
Synthetic biodegradable polymers as orthopedic devices
John C. Middleton*, Arthur J. Tipton
Birmingham Polymers, Inc. 756 Tom Martin Drive, Birmingham, AL 35211, USA
Abstract
Polymer scientists, working closely with those in the device and medical "elds, have made tremendous advances over the past 30
years in the use of synthetic materials in the body. In this article we will focus on properties of biodegradable polymers which make
them ideally suited for orthopedic applications where a permanent implant is not desired. The materials with the greatest history of
use are the poly(lactides) and poly(glycolides), and these will be covered in speci"c detail. The chemistry of the polymers, including
synthesis and degradation, the tailoring of properties by proper synthetic controls such as copolymer composition, special
requirements for processing and handling, and mechanisms of biodegradation will be covered. An overview of biocompatibility and
approved devices of particular interest in orthopedics are also covered.
2000 Elsevier Science Ltd. All rights reserved.
Keywords: Biodegradable; Bioabsorbable; Polylactide (PLA); Polyglycolide (PLG)
1. Introduction
Research in the "rst half of the 20th century with
polymers synthesized from glycolic acid and other hydroxy acids was abandoned for further development
because the resulting polymers were too unstable for
long-term industrial uses. However, this very instability
* leading to biodegradation * has proven to be immensely important in medical uses in the last three decades. Polymers prepared from glycolic acid and lactic
acid have found a multitude of uses in the medical industry, beginning with biodegradable sutures "rst approved
in the 1960 s [1]. Since that time other medical devices,
based on lactic and glycolic acid, as well as other
materials, including poly(dioxanone), poly(trimethylene
carbonate) copolymers, and poly( -caprolactone)
homopolymers and copolymers, have been accepted for
use as medical devices [2]. In addition to these approved
devices, a great deal of research continues on polyanhydrides [3], polyorthoesters [4], and other materials
[5,6].
Why would a medical practitioner want a material to
degrade? There may be a variety of reasons, but the most
basic begins with the physician's simple desire: to have
* Corresponding author. Fax: #1-205-917-2245.
E-mail address: jmiddleton@bpi-sbs.com (J.C. Middleton).
a device, which can be used as an implant and will not
necessitate a second surgical event for removal. In addition to not requiring a second surgery, the biodegradation may o!er other advantages. For example, a
fractured bone, "xated with a rigid, non-biodegradable
stainless steel implant, has a tendency for re-fracture
upon removal of the implant. The bone does not carry
su$cient load during the healing process, because the
load is carried by the rigid stainless steel. However an
implant prepared from biodegradable polymer can be
engineered to degrade at a rate that will slowly transfer
load to the healing bone [7]. Another exciting application for which biodegradable polymers o!er tremendous
potential is the basis for drug delivery, either as a drug
delivery system alone or in conjunction to functioning as
a medical device. In orthopedic applications, the delivery
of a bone morphogenic protein may be used to speed the
healing process after a fracture [8], or the delivery of an
antibiotic may help prevent osteomyelitis following
surgery [9].
Polymer scientists, working closely with those in the
device and medical "elds, have made tremendous advances over the past 30 years. In this article we will focus
on a number of these devices. We will also cover the
chemistry of the polymers, including synthesis and degradation, how properties can be controlled by proper synthetic controls such as copolymer composition, special
requirements for processing and handling, and discuss
some of the commercial devices.
0142-9612/00/$ - see front matter
2000 Elsevier Science Ltd. All rights reserved.
PII: S 0 1 4 2 - 9 6 1 2 ( 0 0 ) 0 0 1 0 1 - 0
Jbmt=1471=Ravi=VVC
2336
J.C. Middleton, A.J. Tipton / Biomaterials 21 (2000) 2335}2346
Nomenclature
Abbreviations
LPLA
poly(L-lactide)
PGA
poly(glycolide)
DLPLA
poly(DL-lactide)
PDO
poly(dioxanone)
LDLPLA
poly(DL-lactide-co-L-lactide)
SR
self-reinforced
DLPLG
poly(DL-lactide-co-glycolide)
PGA-TMC poly(glycolide-co-trimethylene carbonate)
LPLG
poly(L-lactide-co-glycolide)
PCL
poly( -caprolactone)
The general criteria for selecting a polymer for use as
a biomaterial is to match the mechanical properties and
the time of degradation to the needs of the application.
The ideal polymer for an application would have the
following properties:
E
E
E
E
E
does not evoke an in#ammatory/toxic response, disproportionate to its bene"cial e!ect,
is metabolized in the body after ful"lling its purpose
leaving no trace,
is easily processed into the "nal product form,
has acceptable shelf life,
is easily sterilized.
The mechanical properties match the application so
that su$cient strength remains until the surrounding
tissue has healed
2. Synthesis
As expected, biodegradable polymers can be either
natural or synthetic. Here we will cover uses and properties of synthetic biodegradable polymers. These synthetic
polymers in general o!er greater advantages over natural
materials in that they can be tailored to give a wider
range of properties and have more predictable lot-to-lot
uniformity than materials from natural sources. Also
a more reliable source of raw materials is obtained with
synthetic polymers that are free of concerns of immunogenicity [2].
The factors that a!ect the mechanical performance of
biodegradable polymers are those that are well known to
the polymer scientist. These factors are monomer selection, initiator selection, process conditions, and the presence of additives. These factors in turn in#uence the
polymer's hydrophilicity, crystallinity, melt and glass
transition temperatures, molecular weight, molecular
weight distribution, end groups, sequence distribution
(random versus blocky), and the presence of residual
monomer or additives [10]. In addition, the polymer
scientist working with biodegradable polymers must also
evaluate each of these variables for its e!ect on biodegradation. Examples will be given throughout the text
illustrating how some of these variables a!ect performance.
Biodegradation has been accomplished by synthesizing polymers that have hydrolytically unstable linkages
in the backbone. These most common chemical functional groups are esters, anhydrides, orthoesters, and
amides.
The following is an overview of the synthetic-biodegradable polymers that are currently being used or investigated for use as wound closure (sutures, staples), and
orthopedic "xation devices (pins, rods, screws, tacks,
ligaments). Most of the commercially available biodegradable devices are polyesters composed of
homopolymers or copolymers of glycolide and lactide.
There are also products made from copolymers of
trimethylene carbonate, -caprolactone, and polydioxanone.
2.1. Notation
A polymer is generally named based on the monomer
it is synthesized from. For example, ethylene is used to
produce poly(ethylene). For both glycolic acid and lactic
acid, an intermediate cyclic dimer is prepared and puri"ed, prior to polymerization. These dimers are called
glycolide and lactide, respectively. Although most references in the literature refer to poly(glycolide) or
poly(lactide), you will also "nd references to poly(glycolic
acid) and poly(lactic acid). Poly(lactide) exists in two
stereo forms, signi"ed by a D or L for dexorotary or
levorotary, or by DL for the racemic mix.
Poly(glycolide) (PGA) Poly(glycolide) is the simplest
linear aliphatic polyester. PGA was used to develop the
"rst totally synthetic absorbable suture that has been
marketed as DEXON since the 1960s by Davis and
Geck [5,6]. Glycolide monomer is synthesized from the
dimerization of glycolic acid. The ring opening polymerization of glycolide yields high-molecular-weight
materials with about 1}3% residual monomer present
(Fig. 1). PGA is highly crystalline (45}55%) with a high
melting point (220}2253C) and a glass transition temperature of 35}403C [6]. Because of its high degree of crystallization, it is not soluble in most organic solvents; the
Fig. 1. Synthesis of poly(glycolide) (PGA).
Jbmt=1471=Ravi=VVC
J.C. Middleton, A.J. Tipton / Biomaterials 21 (2000) 2335}2346
2337
Table 1
List of commercial biodegradable devices [13]
Application
Trade name
Composition
Manufacturer
Fracture "xation
Fracture "xation
Fracture "xation
Fracture "xation
Fracture "xation
Fracture "xation
Interference screws
Interference screws
Interference screws
Interference screws
Interference screws
Interference screws
Interference screws
Suture anchors
Suture anchors
Suture anchors
Suture anchors
Suture anchors
Suture anchors
Suture anchors
Suture anchors
Suture anchors
Suture anchors
Suture anchors
Suture anchors
Suture anchors
Suture anchors
Suture anchors
Suture anchors
Suture anchors
Suture anchors
Suture anchors
Suture anchors
Suture anchors
Craniomaxillofacial "xation
Meniscus repair
Meniscus repair
Meniscus repair
ACL reconstruction
Meniscus repair
Meniscus repair
SmartPins
SmartPins
SmartScrew
SmartTack
Phantom SofThread Soft Tissue Fixation Screw
Orthosorb Pin
Full Thread Bio}Interference Screw
Sheathed Bio}Interference Screw
Phantom Interference Screw
Biologically Quiet Interference Screw
BioScrew
Sysorb
Endo}Fix Screw
Bankart Tack
SmartAnchor-D
SmartAnchor-L
Phantom Suture Anchor
BioROC EZ 2.8 mm
BioROC EZ 3.5 mm
Biologically Quiet Biosphere
Biologically Quiet Mini-Screw
Bio-Anchor
GLS
Panalok
Panalok RC
Suretak 6.0
Suretak 8.0
Suretak II w spikes
TAG 3.7 mm Wedge
TAG Rod II
SD sorb 2 mm
SD sorb 3mm
SD sorb E}Z TAC
Bio}Statak
LactoSorb Screws and Plates
Menicus Arrow
Clear"x Meniscal Dart
Clear"x Meniscal Screw
Biologically Quiet Staple
Meniscal Stinger
SD sorb Meniscal Staple
SR-LPLA
SR-PGA
SR-LPLA
SR-LPLA
LPLA
PDO
LPLA
LPLA
LPLA
85/15 DLPLG
LPLA
LLPLA
PGA}TMC
SR-LPLA
SR-LPLA
SR-LPLA
LPLA
LPLA
LPLA
85/15 DLPLG
85/15 DLPLG
LPLA
LPLA
LPLA
LPLA
PGA}TMC
PGA}TMC
PGA}TMC
PGA}TMC
PGA}TMC
82/18 LPLG
82/18 LPLG
82/18 LPLG
LPLA
82/18 LPLG
SR}LPLA
LPLA
LPLA
85/15 DLPLG
LPLA
82/18 LPLG
Bionx Implants
Bionx Implants
Bionx Implants
Bionx Implants
DePuy
J & J Orthopedics
Arthrex
Arthrex
DuPuy
Instrument Makar
Linvatec
Sulzer Orthopedics
Smith and Nephew
Bionx Implants
Bionx Implants
Bionx Implants
DuPuy
Innovasive Devices
Innovasive Devices
Instrument Makar
Instrument Makar
Linvatec
Mitek Products
Mitek Products
Mitek Products
Smith and Nephew
Smith and Nephew
Smith and Nephew
Smith and Nephew
Smith and Nephew
Surgical Dynamics
Surgical Dynamics
Surgical Dynamics
Zimmer
Biomet
Bionx Implants
Innovasive Devices
Innovasive Devices
Instrument Makar
Linvatec
Surgical Dynamics
exceptions are highly #uorinated organic solvents such as
hexa#uoroisopropanol. Fibers from PGA exhibit high
strength and modulus and are too sti! to be used as
sutures except as braided material. Sutures of PGA lose
about 50% of their strength after two weeks and 100% at
four weeks and are completely absorbed in 4}6 months
[6]. Glycolide has been copolymerized with other monomers to reduce the sti!ness of the resulting "bers [11,12].
Barber [13] has reviewed the commercially available
orthopedic devices and only one device was made of
PGA (Table 1).
Poly(lactide) (PLA) Lactide is the cyclic dimer of lactic
acid, which exists as two optical isomers, D and L. Llactide, is the naturally occurring isomer, and DL-lactide
is the synthetic blend of D-lactide and L-lactide. The
Fig. 2. Synthesis of poly(lactide) (PLA).
polymerization of lactide is similar to that of glycolide
(Fig. 2). The homopolymer of L-lactide (LPLA) is a semicrystalline polymer. PGA and LPLA exhibit high tensile
strength and low elongation and consequently have
Jbmt=1471=Ravi=VVC
2338
J.C. Middleton, A.J. Tipton / Biomaterials 21 (2000) 2335}2346
a high modulus that makes them more applicable than
the amorphous polymers for load-bearing applications
such as in orthopedic "xation and sutures. Poly(DLlactide) (DLPLA) is an amorphous polymer having
a random distribution of both isomeric forms of lactic
acid and accordingly is unable to arrange into a crystalline organized structure. This material has lower tensile
strength and higher elongation and much more rapid
degradation time making it more attractive as a drug
delivery system. Poly(L-lactide) is about 37% crystalline
with a melting point of 175}1783C and a glass transition
temperature of 60}653C [14,15]. The degradation time of
LPLA is much slower than that of DLPLA requiring
greater than 2 years to be completely absorbed [16].
Copolymers of L-lactide with glycolide or DL-lactide
have been prepared to disrupt the L-lactide crystallinity
accelerating the degradation process [1,6]. Barber's review of 40 commercial orthopedic devices listed 22 of the
devices as being composed of LPLA [13] (Table 1).
Poly( -caprolactone) (PCL): The ring opening polymerization of -caprolactone (Fig. 3) yields a semicrystalline polymer with a melting point of 59}643C and
a glass-transition temperature of !603C. The
homopolymer has a degradation time of the order of two
years. Copolymers of -caprolactone with DL-lactide
have been synthesized to yield materials with more rapid
degradation rates [17]. A block copolymer of -caprolactone with glycolide that has reduced sti!ness compared
to pure PGA is being sold as a mono"lament suture
under the trade name MONOCRYL
by Ethicon
[5,11,12], but no commercial medical devices are listed
by Barber as made of PCL [13].
Poly(dioxanone) (a polyether-ester): The ring opening
polymerization of p-dioxanone resulted in the "rst clinically tested mono"lament synthetic suture that is known
as PDS marketed by Ethicon (Fig. 4). This material has
about 55% crystallinity with a glass-transition temperature of !10 to 03C. Poly(dioxanone) demonstrated no
acute or toxic e!ects on implantation [6]. Johnson and
Johnson Orthopedics has an absorbable pin for fracture
"xation composed of poly(dioxanone) on the market
[13].
Poly(lactide-co-glycolide) (PLG): Using the polyglycolide and poly(L-lactide) properties as base materials, it is
possible to copolymerize the two monomers to extend
the range of homopolymer properties (Fig. 5).
Copolymers of glycolide with both L-lactide and DLlactide have been developed for both device and drugdelivery applications. It is important to note that there is
not a linear relationship between the copolymer composition and the mechanical and degradation properties
of the materials. For example, a copolymer of 50%
glycolide and 50% DL-lactide degrades faster than either
homopolymer (Fig. 6) [18]. Copolymers of L-lactide with
Fig. 3. Synthesis of poly( -caprolactone) (PCL).
Fig. 4. Synthesis of poly(dioxanone) (PDS).
Fig. 6. Half-life of PLA and PGA homopolymers and copolymers
implanted in rat tissue [11].
Fig. 5. Synthesis of poly(lactide-co-glycolide) (PLG).
Jbmt=1471=Ravi=VVC
J.C. Middleton, A.J. Tipton / Biomaterials 21 (2000) 2335}2346
2339
Fig. 7. Synthesis of poly(glycolide-co-trimethylene carbonate) (PGA}TMC).
25}70% glycolide are amorphous due to the disruption
of the regularity of the polymer chain by the other monomer [1]. The Biologically Quiet2+ line of products by
Instrument Makar are composed of an 85/15 poly(DLlactide-co-glycolide). Surgical Dynamics and Biomet
have chosen an 82/18 poly(L-lactide-co-glycolide)
copolymer for use as suture anchors and as screws
and plates for craniomaxillofacial repair respectively
[13, 19].
Copolymers of glycolide with trimethylene carbonate
(TMC) called polyglyconate have been prepared as both
sutures (MAXON , Davis and Geck) [12] and as tacks
and screws (Smith and Nephew Endoscopy) [13]. Typically these are prepared as A}B}A block copolymers in
a 2 : 1 glycolide : TMC ratio with a glycolide-TMC center
block (B) and pure glycolide end blocks (A) (Fig. 7). These
materials have better #exibility than pure PGA and are
absorbed in about seven months [6]. Glycolide has also
been polymerized with TMC and p-dioxanone (BIOSYN by US Surgical) to form a terpolymer suture with
reduced sti!ness compared to pure PGA "bers, with
absorption within 3}4 months [13].
Currently, only resorbable "xation devices made from
homopolymers or copolymers of glycolide, lactide, caprolactone, p-dioxanone and trimethylene carbonate have
been commercialized [13]. There are other polymers,
however, that are being investigated for use as materials
for biodegradable devices that merit mentioning.
Poly(amino acids): The use of synthetic poly(amino
acids) as polymers for biomedical devices would seem
a logical choice because of their wide occurrence in
nature. However, in practice, pure insoluble poly(amino
acids) have found little utility due to their high crystallinity which makes them di!cult to process and gives
relatively slow degradation. Also, the antigenicity of
polymers with more than three amino acids in the chain
also makes them inappropriate for use in vivo [20]. To
circumvent these problems, modi"ed `pseudoa
poly(amino acids) have been synthesized using a tyrosine
derivative. Tyrosine-derived polycarbonates are high
strength materials that may be useful as orthopedic
implants [5,20].
The search for new candidate polymers for drug delivery may o!er potential for medical device applications as
Fig. 8. Molecular structure of a polyanhydride.
Fig. 9. Molecular structure of a polyorthoester.
well. In drug delivery the formulation scientist is concerned not only with shelf life stability of the drug but
also with stability after implantation, where the drug may
reside in the implant for 1}6 months or more. For drugs
that are hydrolytically unstable, a polymer that absorbs
water may be counter-indicated, so researchers began
evaluating more hydrophobic polymers that degrade by
surface erosion rather than bulk hydrolytic degradation.
Two classes of these polymers are the polyanhydrides
and the polyorthoesters.
Polyanhydrides: Polyanhydrides have been synthesized
by the dehydration of diacid molecules by melt polycondensation (Fig. 8). Degradation times may be adjusted
from days to years by degree of hydrophobicity of monomer selection. They degrade primarily by surface erosion
and possess excellent in vivo compatibility. So far they
have been only approved as a drug delivery system. The
Gliadel product designed for delivery of BCNU in the
brain was approved by the FDA in 1996 and is being
produced by Guilford [3,5].
Polyorthoesters: Polyorthoesters were "rst investigated
in the 1970s by the Alza Corporation and SRI International in search of a new synthetic biodegradable
polymer for drug-delivery applications (Fig. 9). These
Jbmt=1471=Ravi=VVC
2340
J.C. Middleton, A.J. Tipton / Biomaterials 21 (2000) 2335}2346
materials have gone through several generations of synthetic improvements to yield materials that can be polymerized at room temperature without production of
condensation by-products. These materials are hydrophobic with hydrolytic linkages that are acid-sensitive,
but stable to base. They degrade by surface erosion and
degradation rates may be controlled by incorporation of
acidic or basic excipients [2,4,5].
3. Physical properties
The selection of a material for an orthopedic implant
depends on the mechanical properties needed for the
application and the degradation time desired. Polymers
may be either semicrystalline or amorphous. Semicrystalline polymers have regular repeating units that allow the
chains to fold into dense regions called crystallites. These
act as crosslinks giving the polymer higher tensile
strengths and higher modulus (sti!ness) as compared to
an amorphous analog. No polymer can completely organize into a fully crystalline material so there are still
amorphous areas in semicrystalline polymers. When
a semicrystalline polymer is raised above its melting
point (¹ ) it may be shaped into rods or molded parts.
Amorphous polymers and the amorphous regions of
semicrystalline polymers exhibit a glass transition temperature or ¹ . At temperatures above ¹ , a polymer
acts more like a rubber and at temperatures below ¹ ,
a polymer acts more like a glass. A polymer that has
a ¹ around body temperature may be much more ductile when implanted than it appears to be at room temperature. These properties can a!ect both the mechanical
properties as well as the degradation time of the implant
[10,21]. For the polyesters, the presence of water can act
as a plasticizer and lower the ¹ and a!ect degradation
and mechanical properties. Koelling et al. [22] evaluated
the mechanical properties of 90/10 poly(L-lactide-coDL-lactide) under both wet and dry conditions. They saw
the mechanical properties were lower for the polymers
tested in the wet condition.
A good example of the di!erences between a semicrystalline and amorphous polymer is illustrated by the
di!erences between poly(L-lactide) and poly(DL-lactide)
discussed earlier under the synthesis section. The semicrystalline poly(L-lactide) has a modulus about 25% higher than poly(DL-lactide) and a degradation time on the
order of 3 to 5 years. The amorphous poly(DL-lactide) has
a degradation time of 12 to 16 months [21,23,24].
A common way of a!ecting crystallinity is by the use of
comonomers in the synthesis. Unlike monomers do not
typically co-crystallize and crystallinity can be disrupted
by copolymerization, with the e!ect being more pronounced at higher comonomer levels. For example, both
glycolide and L-lactide homopolymers are semicrystalline, and copolymers of L-lactide and glycolide exhibit
some crystallinity when either monomer is present over
70 mol% [1]. Copolymers of DL-lactide and glycolide
are amorphous when DL-lactide is the major component
[23]. For applications where an implant will be under
substantial load the family of semicrystalline biodegradable polymers would typically be chosen. Daniels et al.
[14] have reviewed the mechanical properties for both
reinforced and unreinforced biodegradable polymers.
Table 2 shows some of the physical properties and degradation times for selected biodegradable polymers.
4. Processing
Biodegradable polymers may be processed similar to
any engineering thermoplastic in that they can be melted
Table 2
Physical, mechanical, and degradation properties of selected biodegradable polymers; bone and steel included as reference materials [20,21,23]
Polymer
Melting point (3C)
Glass transition
temperature (3C)
Modulus (Gpa)
Elongation (%)
Degradation time
(months)
PGA
LPLA
DLPLA
PCl
PDO
PGA}TMC
85/15 DLPLG
75/25 DLPLG
65/35 DLPLG
50/50 DLPLG
Bone
Steel
225}230
173}178
Amorphous
58}63
N/A
N/A
Amorphous
Amorphous
Amorphous
Amorphous
35}40
60}65
55}60
!65}!60
!10}0
N/A
50}55
50}55
45}50
45}50
7.0
2.7
1.9
0.4
1.5
2.4
2.0
2.0
2.0
2.0
10}20
210
15}20
5}10
3}10
300}500
N/A
N/A
3}10
3}10
3}10
3}10
6 to 12
'24
12 to 16
'24
6 to 12
6 to 12
5 to 6
4 to 5
3 to 4
1 to 2
Tensile or #exural modulus.
Time to complete resorption.
Jbmt=1471=Ravi=VVC
J.C. Middleton, A.J. Tipton / Biomaterials 21 (2000) 2335}2346
and formed into "bers, rods and molded parts. Final
parts can be extruded, injection molded, compression
molded, or solvent spun or cast. In some circumstances
the primary processing may be followed by subsequent
machining into "nal parts.
The additional complication during processing is the
potential for molecular weight decrease due to the hydrolytic sensitivity of the polymer bonds. The presence of
moisture during processing can reduce the molecular
weight and alter the "nal polymer properties. To avoid
hydrolytic degradation during processing, extra precautions need to be taken to dry the polymer before
thermally processing and preventing moisture from contacting the polymer during processing. Michaeli and von
Oepen [25,26] have studied the in#uence of several processing factors on degradation during processing. Drying
a polymer 24 h at 803C prior to processing reduced
degradation by approximately 30% when processing
above 2003C. Drying may be accomplished by vacuum
drying or drying in a resorption circulating air dryer.
Von Oepen reported drying semicrystalline polymers at
1403C resulted in moisture contents of less than 0.02%
without incurring degradation during drying. They recommend moisture content not to exceed 0.02% to avoid
excessive degradation during processing [25]. Michaeli
and von Oepen reported that most of the moisture is
removed after 4 h drying [26]. Middleton et al. [27]
reported the e!ects of drying on the melt viscosity of
PGA when processed at 2503C. Here the polymer was
vacuum dried 24 h at room temperature followed by
vacuum drying 24 h at 1003C. This drying cycle reduced
the moisture from 0.02 to 0.003%. PGA processed at
2503C with 0.02% moisture resulted in over 50% degradation as indicated by a decrease in melt viscosity, whereas drying to 0.003% did not. Care must be exercised
when drying polymers above room temperature. For
example, amorphous polymer pellets may fuse when the
drying temperature exceeds the glass transition temperature. Most of the amorphous polymers should only be
dried at room temperature.
Other techniques may also be used to prevent moisture
from entering the fabrication process. Packaging the
polymers in small quantities is recommended so that
the material is used up quickly during processing once
the package is opened to prevent moisture absorption
over time. Blanketing the material hopper or material
inlet with nitrogen or dried air will also prevent moisture
from entering the system.
Most synthetic, resorbable polymers have been synthesized by ring-opening polymerization and there exists
a thermodynamic equilibrium between the polymerization reaction and the reverse reaction that will result in
monomer formation. Excessively high processing temperatures can push the equilibrium to depolymerization
resulting in monomer formation during the molding or
extrusion process. The presence of excess monomer may
2341
act as a plasticizer changing the mechanical properties
and may catalyze the hydrolysis of the device resulting in
altered degradation kinetics [6].
There are also strong interactions among temperature,
moisture content, shear rate, and residence time in the
machine. Residence time is de"ned as time at temperature the material is in the barrel of a molding machine.
Michaeli and von Oepen [25,26] have studied the e!ect
of these interactions on polymer degradation for LPLA.
When the temperature was raised from 190 to 2303C all
the other e!ects were inconsequential. Higher shear rates
and longer residence times resulted in increasing polymer
degradation even at lower temperatures. In general, processing at the mildest conditions possible and the rigorous exclusion of moisture are the recommended. In many
cases this is di$cult as the devices being extruded or
molded are small "bers or parts from very high-molecular-weight polymer. High temperatures are often needed
to reduce the melt viscosity or high pressures needed to
enable the polymer to #ow through small ori"ces to
create "ber or "ll a mold. Several iterations of molding or
extrusion may be needed to get the "nal part properties
necessary for the application.
5. Packaging and sterilization
Because these polymers are hydrolytically unstable,
the presence of moisture can degrade them in storage,
during processing (as already discussed), and after device
fabrication. The solution for hydrolysis instability is
simple in theory: eliminate the moisture and eliminate the
degradation. Because the materials are naturally hygroscopic, eliminating water and keeping the polymer free of
water are di$cult. The as-synthesized polymers have
relatively low water contents, as any residual water in the
monomer is consumed in the polymerization reaction.
The polymers are quickly packaged after manufacture
and generally double bagged under an inert atmosphere
or vacuum. The bag material may be polymeric or foil,
but it must be very resistant to water permeability [23].
The polymers are typically stored in a freezer to minimize
the e!ects of moisture present. The packaged polymer
should always be at room temperature when opened to
minimize condensation and should be handled as little as
possible at ambient atmospheric conditions [5]. As expected, there is a relation between biodegradation rate,
shelf stability and polymer properties. For example, the
more hydrophilic glycolide polymers are much more
sensitive to hydrolytic degradation than those prepared
from the more hydrophobic lactide. Williams et al. [28]
have studied six di!erent storage conditions for biodegradable polymers and found that the polymers remain
stable even at room temperature for over two years as
indicated by molecular weight retention when packaged
in desiccated moisture proof bags.
Jbmt=1471=Ravi=VVC
2342
J.C. Middleton, A.J. Tipton / Biomaterials 21 (2000) 2335}2346
Final packaging consists of placing the suture or device in an airtight moisture-proof container. A desiccant
can be added to reduce the e!ect of moisture. For
example, sutures are wrapped around a specially dried
paper holder that acts as a desiccant. In some cases the
"nal device may be stored at sub-ambient temperature as
an added precaution against degradation.
The "nal devices should not be sterilized by autoclaving or dry heat because this will degrade the device.
Typically the device is sterilized by -radiation, ethylene
oxide (EtO), or other less-known techniques such as
plasma etching [5,7,25] or electron beam irradiation.
Both radiation and EtO have disadvantages. Radiation,
particularly at doses above 2 Mrad, can result in signi"cant degradation of the polymer chain, resulting in
reduced molecular weight and in#uencing "nal mechanical properties and degradation times [6,8,15].
Poly(glycolide), poly(lactide) and poly(dioxanone) are especially sensitive to -radiation, and these are usually
sterilized for device applications by exposure to ethylene
oxide. The use of highly toxic EtO presents a safety
hazard, so great care is used to ensure that all the gas is
removed from the device before "nal packaging [5]. This
may result in extremely long vacuum aeration times. One
researcher has recommended a period of over 2 weeks
[29] to fully remove the residual EtO gas. The temperature and humidity conditions should also be considered
when submitting devices for sterilization. Temperatures
must be kept below the glass transition temperature of
the polymer to prevent the part geometry from changing
during sterilization. If necessary, parts can be kept at 03C
or lower during the irradiation process. Of the 40 commercial orthopedic devices listed in Barber's review 25
were sterilized by EtO and 15 by irradiation [13]. No
other techniques were listed.
6. Degradation
Once implanted in the body, the biodegradable device
should maintain mechanical properties until it is no
longer needed and then be degraded, absorbed, and excreted by the body, leaving no trace. Simple chemical
hydrolysis of the hydrolytically unstable backbone is the
prevailing mechanism for the polymer degradation. For
semicrystalline polymers this occurs in two phases. In the
"rst phase, water penetrates the bulk of the device, preferentially attacking the chemical bonds in the amorphous
phase and converting long polymer chains into shorter,
ultimately water-soluble fragments. Because this occurs
in the amorphous phase initially there is a reduction in
molecular weight without a loss in physical properties as
the device matrix is still held together by the crystalline
regions. The reduction in molecular weight is soon followed by a reduction in physical properties as water
begins to fragment the device. In the second phase, enzy-
Fig. 10. Generic curves showing the sequence of polymer-molecular
weight, strength, and mass-reduction over time [19].
matic attack of the fragments occurs. The metabolizing
of the fragments results in a rapid loss of polymer mass
(Fig. 10) [21].
Bulk erosion occurs when the rate at which water
penetrates the device exceeds that at which the polymer is
converted into water-soluble materials (resulting in erosion throughout the device). The lactide and glycolide
commercially available devices and sutures degrade by
bulk erosion [5]. This two-stage degradation mechanism
has led one researcher to report that the degradation rate
at the surface of large lactide}glycolide implants is slower
that the degradation in the interior [30]. Initially, degradation does occur more rapidly at the surface due to the
greater availability of water. The degradation products at
the surface are rapidly dissolved in the surrounding #uid
and removed from the bulk polymer. In the interior of the
device the inability of large polymeric degradation products to di!use away from the bulk device results in a local
acidic environment in the interior of the implant. The
increased acidic environment catalyses further degradation resulting in accelerated hydrolysis of the ester
linkages in the interior. Athanasiou [31] has shown that
low-porosity implants from 50/50 DLPLG degrade faster than high-porosity implants. He attributes this to the
quick di!usion of low pH degradants from the interior of
the high-porosity devices.
Polymer scientists have used this knowledge to tailor
the degradation rates of biodegradable polymers. Tracy
[32] reported the e!ects of replacing ester end groups
with carboxylic acid end groups on DLPLG polymers
(Figs. 11 and 12) accelerated both water uptake and
degradation rate in vitro. The acid-end groups both add
to the hydrophilicity of the polymer and catalyze degradation. Middleton et al. [33] conducted an in vitro degradation study in phosphate-bu!ered saline comparing
Jbmt=1471=Ravi=VVC
J.C. Middleton, A.J. Tipton / Biomaterials 21 (2000) 2335}2346
2343
Fig. 11. Initiation of DL-lactide with an alcohol resulting in a polymer with one ester end group and one alcohol end group.
Fig. 12. Initiation of DL-lactide with water resulting in a polymer with one carboxylic-acid end group and one alcohol-end group.
Fig. 13. Initiation of DL-lactide and glycolide with a monofunctional poly(ethyleneglycol) resulting a block copolymer consisting of a PEG block and
a poly(DL-lactide-co-glycolide) block.
DLPLG rods where the ester end groups were
replaced with covalently bound monofunctional poly(ethylene glycol) (mPEG) (Fig. 13). The mPEG-DLPLG
demonstrated enhanced water uptake without accelerated degradation. It is believed the presence of the ethylene glycol units enhanced polymer hydrophilicity
without lowering the pH of the local environment. By
increasing the water uptake it may also have allowed the
acidic degradants to more readily di!use away from the
interior of the rod.
There is a second type of biodegradation called surface
erosion when the rate at which the polymer penetrates
the device is slower than the rate of conversion of the
polymer into water-soluble materials [5]. Surface erosion
results in the device thinning over time while maintaining
its bulk integrity. Polyanhydrides and polyorthoesters
are examples of this type of erosion when the polymer is
hydrophobic, but the chemical bonds are highly susceptible to hydrolysis. In general, this process is referred to in
the literature as bioerosion rather than biodegradation.
The degradation-absorption mechanism is the result of
many interrelated factors, including:
E
E
E
the chemical stability of the polymer backbone,
the presence of catalysts,
additives,
Jbmt=1471=Ravi=VVC
2344
E
E
E
J.C. Middleton, A.J. Tipton / Biomaterials 21 (2000) 2335}2346
impurities or plasticizers,
the geometry of the device,
the location of the device.
The balancing of these factors to tailor an implant to
slowly degrade and transfer stress to the surrounding
tissue as it heals at the appropriate rate is one of the
major challenges facing the researchers today.
The factors which accelerate polymer degradation are
the following
E
E
E
E
E
More hydrophilic monomer.
More hydrophilic, acidic endgroups.
More reactive hydrolytic group in the backbone.
Less crystallinity.
Smaller device size.
The location of the device can play an important role
in the degradation rate of lactide}glycolide implants.
Large devices implanted in areas with poor vascularization may degrade and overwhelm the body's ability to
#ush away degradants. This leads to a build up of acidic
by-products. An acidic environment will catalyze the
further degradation and cause further reduction in pH
[7]. The local reduction in pH may also be responsible
for adverse tissue reactions [34]. It has also been reported [35] that implants under stress degrade faster. It
was proposed that the stressed implant may form microcracks increasing the surface area exposed to water [7].
They removed and analyzed the remaining LPLA material after 3.3 to 5.7 years. Some highly crystalline LPLA
particles remained after 5.7 years. The particles were not
irritable and did not cause injury to the cell, but did
induce a reaction in the form of detectable swelling.
Barber [36] evaluated 85 patients in two groups that
received either a metal or LPLA interference screw. No
statistical di!erences were found between the two groups
after two years.
As with the other areas we have explored, there are
many factors that may in#uence the reaction of the body
to the presence of a biodegradable implant. The response
may be related to the size and composition of the implant
as well as the implant site. The degradation rate of the
polymer and the implant site's ability to eliminate the
acidic degradants play an important role in the local
tissue's reaction to the implant. If the surrounding tissue
cannot eliminate the acid by-products of a rapidly degrading implant then an in#ammatory or toxic response
may result [34].
In summary, the results from studies in humans are
mostly favorable with some negative reports. The complications arising from biodegradable orthopedic implants of polymers of lactide and glycolide typically
occur at a rate of less than 10% [7]. Although initially
signi"cant, these problems resolve with time making the
future of biodegradable implants bright.
8. Commercial biodegradable devices
7. Biocompatibility
Some of the requirements listed in the earlier section of
this article stated that the ideal implant would not invoke
an in#ammatory or toxic response and that the degradants must be metabolized in the body after ful"lling its
purpose leaving no trace. For example, poly(lactide) hydrolyzes to lactic acid that is a normal product of muscular contraction in animals. The lactic acid is then further
metabolized through the tricarboxylic acid cycle and
then excreted as carbon dioxide and water. Poly(glycolide) is degraded by hydrolysis and esterases to glycolic
acid. Glycolic acid monomer may be excreted directly in
urine or may react to form glycine. Glycine can be used
to synthesize serine and subsequently transformed
into pyruvic acid where it enters the tricarboxylic acid
cycle [7].
There have been numerous studies on the biocompatibility of implants since the early 1960s mostly focusing on
polymers of lactide and glycolide. The majority of results
indicate that these polymers are su$ciently biocompatible, with a minority suggesting otherwise. The literature
before 1993 has been summarized by Agrawal et al. [15].
Bergsma et al. [16] conducted a study on patients that
have received PLLA implants for zygomatic fractures.
The total US revenues from commercial products developed from absorbable polymers in 1995 was estimated
to be over $300 million with over 95% of revenues
generated from the sale of bioabsorbable sutures. The
other 5% is attributed to orthopedic "xation devices in
the forms of pins, rods and tacks, staples for wound
closure, and dental applications [37]. In addition, research into biodegradable systems continues to increase,
with 60 to 70 papers published each year in the late 1970s
to over 400 each year in the early 1990s. The rate at
which bioabsorbable "xation devices are cleared through
the 510(k) process by the US Food and Drug Administration (FDA) is also increasing, with seven approved in
1995 [19].
The two routes for getting FDA approval to market
and sell medical devices in the US are the 510(k) process
and the premarket approval (PMA) process. The 510(k)
process requires that the new device be shown to be
equivalent to a device currently on the US market in
terms of safety and e$cacy. Clinical data may or may not
be required for these devices. The PMA process always
requires clinical data and is as stringent as the requirements for a new drug application.
Orthopedic "xation devices of synthetic biodegradable
polymers have advantages over metal implants in that
Jbmt=1471=Ravi=VVC
J.C. Middleton, A.J. Tipton / Biomaterials 21 (2000) 2335}2346
they transfer stress over time to the damaged area, allowing healing of the tissues, and eliminate the need for
a subsequent operation for implant removal. The current
approved materials have not been commercialized as
bone plates for long bone support such as the femur.
They have found applications where lower-strength materials are su$cient, such as in the ankle, knee, and hand
areas as interference screws, tacks and pins for ligament
attachment and meniscal repair, suture anchors, and rods
and pins for fracture "xation. Barber has recently compiled a review of the commercially available orthopedic
devices [13]. He grouped the devices into four categories
with the number of devices listed in each category in
parentheses: fracture "xation (6), interference-"xation
screws (6), suture anchors (21) and other devices (7).
Other devices include screws and plates for maxillofacial
repair, tacks for meniscal repair and an implant for ACL
reconstruction. The list appears in Table 1.
In this article we have attempted to provide an overview of the orthopedic uses of biodegradable polymers.
While sutures were the "rst commercial product and still
account for 95% of all sales, a number of products are
now approved for a wide range of applications. And it is
expected that a number of additional products will be
approved in the next decade.
What is it about these materials that makes them so
attractive to the device industry? First, in this conservative "eld, where devices serve critical, perhaps life and
death functions, the industry is slow to accept new materials or new designs. But polymers prepared from these
materials, particularly lactide and glycolide, have a long
history of safety including the approval of several products. Building on this solid history, researchers continue
to evaluate these materials for other uses. The wide range
of properties that can be obtained in polymers built with
these few monomer units has allowed for a variety of
products. We expect that in the future, even more than
today, surgeons will have available a number of products
using biodegradable products that will speed patient
recovery and eliminate follow-up surgeries.
Acknowledgements
The authors would like to thank Ms. Arlene Koelling,
Senior Research Engineer at Smith and Nephew for her
help in preparation of this manuscript.
References
[1] Gilding DK, Reed AM. Biodegradable polymers for use in surgery * polyglycolic/poly(lactic acid) homo * and copolymers: 1.
Polymer 1979;20:1459}64.
[2] Barrows TH. Degradable implant materials: a review of synthetic
absorbable polymers and their applications. Clin Mater
1986;1:233}57.
2345
[3] Domb AJ, Amselem S, Langer R, Maniar M. Polyanhydrides as
carriers of drugs. In: Shalaby SW, editor. Biomedical polymers
Designed to degrade systems. New York: Hanser, 1994. p. 69}96.
[4] Heller J, Daniels AU. Poly(orthoesters). In: Shalaby SW, editor.
Biomedical polymers. Designed to degrade systems. New York:
Hanser, 1994. p. 35}68.
[5] Kohn J, Langer R. Bioresorbable and bioerodible materials. In:
Ratner BD, Ho!man AS, Schoen FJ, Lemons JE, editors. Biomaterials science. New York: Academic Press, 1996. p. 64}72.
[6] Shalaby SW, Johnson RA. Synthetic absorbable polyesters. In:
Shalaby SW, editor. Biomedical polymers. Designed to degrade
systems. New York: Hanser, 1994. p. 1}34.
[7] Athanasiou KA, Agrawal CE, Barber FA, Burkhart SS. Orthopaedic applications for PLA}PGA biodegradable polymers. Arthrosc: J Arthrosc Relat Surg 1998;14(7):726}37.
[8] Wang EA, Rosen V, D'jAlessandro JS, Bauduy M, Cordes P,
Harada T, Isreal DI, Hewick RM, Kerns KM, LaPan P, Luxenberg D, McQuaid D, Moutsatsos IK, Nove J, Wozney JM.
Recombinant human bone morphogenic protein induces bone
formation. Proc Natl Acad Sci 1990;87:2220}4.
[9] Ramchandani M, Robinson D. In vitro release of cipro#oxacin
from PLGA 50 : 50 implants. J Controlled Rel 1998;54:167}75.
[10] Odian G. Principles of polymerization, (2nd ed). New York: Wiley
Interscience, 1981.
[11] Goupil D. Sutures. In: Ratner BD, Ho!man AS, Schoen FJ,
Lemons JE, editors. Biomaterials science. New York: Academic
Press, 1996. p. 356}60.
[12] Lewis OG, Fabisial W. Sutures. In Kirk-Othmer encyclopedia of
chemical technology (4th ed.). New York: Wiley, 1997.
[13] Barber FA. Resorbable "xation devices: a product guide (Orthopedic Special Edition) 1998;4:1111}17.
[14] Daniels AU, Chang MKO, Andriano KP, Heller J. Mechanical
properties of biodegradable polymers and composites proposed
for internal "xation of bone. J Appl Biomater 1990;1:57}78.
[15] Agrawal CM, Niederauer GG, Micallef DM, Athanasiou KA.
The use of PLA}PGA polymers in orthopedics. In: Wise DL,
Trantolo DJ, Altobelli DE, Yaszemski MJ, Greser JD, Schwartz
ER. editors. Encyclopedic handbook of biomaterials and bioengineering Part A. Materials, vol. 2. New York: Marcel Dekker,
1995. p. 1055}89.
[16] Bergsma JE, de Bruijn WC, Rozema FR, Bos RRM, Boering G.
Late degradation tissue response to poly(L-lactide) bone plates
and screws. Biomaterials 1995;16(1):25}31.
[17] Schindler A, Je!coat R, Kimmel GL, Pitt CG, Wall ME, Zwiedinger R. Biodegradable polymers for sustained drug delivery. Contemp Topics Polym Sci 1977;2:251}89.
[18] Miller RA, Brady JM, Cutright DE. Degradation rates of oral
resorbable implants (polylactates and polyglycolates): rate modi"cation with changes in PLA/PGA copolymer ratios. J Biomed
Mater Res 1977;11:711}9.
[19] Pietrzak WS, Verstynen BS, Sarver DR. Bioabsorbable "xation
devices: status for the Craniomaxillofaxial surgeon. J. Craniofaxial Surg 1997;2:92}6.
[20] Nathan A, Kohn J. Amino acid derived polymers. In: Shalaby
SW, editor. Biomedical polymers Designed to degrade systems.
New York: Hanser, 1994. p. 117}51.
[21] Pietrzak WS, Sarver DR, Verstynen BS. Bioabsorbable polymer
science for the practicing surgeon. J. Craniofaxial Surg
1997;2:87}91.
[22] Koelling AS, Ballintyn, NJ, Salehi A, Darden DJ, Taylor, ME,
Varnavas, J, Melton DR. In vitro real-time aging and characterization of poly(L/D,L-lactic acid). In: Bumgardner JD, Puckett
AD, editors. Proceedings of the Sixteenth Biomedical Engineering
Conference. 1997. p. 197}201.
[23] Birmingham Polymers, Inc. Corporate Literature Brochure
C 11-001-B 1999.
[24] Alkermes Inc. Product Literature 1999.
Jbmt=1471=Ravi=VVC
2346
J.C. Middleton, A.J. Tipton / Biomaterials 21 (2000) 2335}2346
[25] Michaeli W, Von Oepen R. Processing of degradable polymers.
ANTEC 1994; 796}804.
[26] Von Oepen R, Michaeli W. Injection moulding of biodegradable
implants. Clin Mater 1992;10:21}8.
[27] Middleton JC, Williams CT, Swaim RP. The melt viscosity of
polyglycolic acid as a function of shear rate, moisture, and inherent viscosity. Trans Soc Biomater 23rd Annu Meeting
1997;20:106.
[28] Williams CT, Middleton JC, Sims KR, Swaim RP, Whit"eld DR,
Yarbrough JC. Long-term stability of biodegradable polymers.
Proceedings of the 17th Southern Biomedical Engineering Conference, 1998 p. 69.
[29] Zislis T, Martin SA, Cerbas E, Heath JR, Mans"eld JL, Hollinger
JO. A scanning electron microscopic study of in vitro toxicity of
ethylene-oxide sterlized bone repair materials. J Oral Impants
1989;25:41}6.
[30] Therin M, Christel P, Li S, Garreau H, Vert M. In vivo degradation of massive poly(alpha hydroxy acids): validation of in vitro
"ndings. Biomaterials 1992;13:594}600.
[31] Athanasiou KA, Schmitz JB, Agrawal CM. The e!ects of porosity
on degradation of PLA-PGA implants. Tissue Engng 1998;4:53}63.
[32] Tracy MA, Firouzabadian L, Zhang Y. E!ects of PLGA end
groups on degradation. Proceedings of the International Symposium on Controlled and Related Bioactive Materials, vol. 22,
1995. p. 786}7.
[33] Middleton JC, Yarbrough JC. The e!ect of PEG end groups on
the degradation of a 75/25 poly(DL-lactide-co-glycolide). Trans
Soc Biomater 25th Annu Meeting 1999;22:535.
[34] Suganuma J, Alexandar H. Biological response of intramedullary bone to poly(L-lactic acid). J Appl Biomater
1993;4:13}27.
[35] Bos RRM, Rozema FR, Boering G, Nijenhuis AJ, Pennings AJ,
Jansen HWB. Bone-plates and screws of bioaborbable poly(Llactide) } an animal pilot study. Br J Oral Maxillofacial Surg
1989;27:467}76.
[36] Barber FA, Burton FE, McGuire DA, Paulos LE. Preliminary
results of an absorbable interference screw. Arthrosc J Arthrosc
Relat Surg 1995;11(5):537}47.
[37] Frost and Sullivan Report. US absorbable and erodible biomaterials products markets; Forecasts of the US absorbable
polymer medical products market. Mountain view, CA: Frost and
Sullivan, 1995 (Chapter 10).